Message
Possible causes
This message is displayed if the selected resolution is so high that less
than 0.8 internal counts per scale interval (d) and/or 0.5 µV/e are avai-
lable if legal-for-trade acc. to OIML/NSC.
This message displays if the maximum load is not an integer multiple of
the scale interval.
Weight units don't match, e.g., subsequent change for [Max] from kg in-
to lb.
This message is displayed when the dead load entered in mV/V plus
scale interval in mV/V is higher than 3 mV/V (= 36 mV).
This message displays, if the scale is not stable.
Remedy
-
Check the mechanical function of the scale.
-
Adapt the filter setting; reduce the resolution.
-
Adapt the standstill conditions.
This message is displayed when the measurement signal is negative
when determining the dead load with [by load].
Cause
Load cell connected with wrong polarity, or defective.
This message is displayed when the dead load entered is
5 mV/V.
16 Error messages
X3 Process Indicator PR 5410