8
Maintenance/repairs/cleaning
8.1
Repairs
Immediately disconnect a defective device from the power supply and notify
Minebea Intec Customer Service.
Defective or damaged cables or screw connections must be replaced as a complete unit.
The seals placed on the device indicate that the device may only be opened and serviced
by authorized specialists to ensure trouble-free and safe operation of the device, and to
ensure that the warranty remains valid.
WARNING
Improper repairs can pose considerable risks to the user.
Repairs should only be performed by Minebea Intec authorized technical personnel
using original spare parts.
8.2
Cleaning
8.2.1
Cleaning instructions
The device should be cleaned at regular intervals.
Disconnect the device from the supply voltage before cleaning.
Condensation may form in the device if it is cleaned with water that is too hot or too cold,
due to temperature differences. Condensation may cause the device to malfunction.
Cleaning the surfaces
All stainless steel parts should be cleaned at regular intervals.
Use a damp cloth or sponge to clean stainless steel parts on the scale (if present). You can
use any commercially available household cleaning agent that is suitable for use on
stainless steel.
Remove all traces of corrosive substances on a regular basis.
8.2.2
Cleaning agents
NOTICE
Some cleaning agents may not be compatible with the device material.
Do not use cleaning agents for stainless steel parts that contain chlorine, alkalines,
acetic acid, hydrochloric acid, sulfuric acid, or citric acid.
The use of cleaning sponges made of steel wool is prohibited (e.g. S.O.S pads).
8 Maintenance/repairs/cleaning