Safety Summary

All safety related regulations, local codes and instructions that

appear in the manual or on equipment must be observed to ensure
personal safety and to prevent damage to either the instrument or
equipment connected to it. If equipment is used in a manner not
specified by the manufacturer, the protection provided by the
equipment may be impaired.

Do not use this unit to directly command motors, valves, or

other actuators not equipped with safeguards. To do so can be
potentially harmful to persons or equipment in the event of a fault
to the unit.

FRONT PANEL

JUMPER LINK FUNCTIONS

The indicator has three jumper links that must be set prior to operation.

These are:

•  

Input Range Select

Select either 

±

24 mVDC or 

±

240 mVDC range. Select an input range

high enough to avoid input signal overload. Program the input range to
match jumper selection.

•  

Bridge Excitation Select 

Select either 5 VDC or 10 VDC bridge excitation.

•  

User Input Source or Sink Select

FRONT

KEY FUNCTIONS

TARA KEY FUNCTION 

The 

TARA

key selects among the Gross, Net and Tare weight displays.

Corresponding panel enunciators illuminate to indicate the selected display. If
Tare weight = 0, the indicator prevents selection of Gross and Net weight
displays. Within the program menu, this key quits programming and returns to
display mode.

PAR KEY FUNCTION

The 

PAR

key allows access to the configuration parameters. The function of

this key can be locked.

Within the program menu, this key stores the active parameter and indexes to

the next.

During Digital Tare, this key stores the tare weight value keyed in.

KEY

The

(Tare In/Tare Out) key both acquires and clears the Tare weight

value. The function of this key can be changed by setting the F1 function
parameter. For data entry, this key is also used to increment a parameter value.

KEY

The

key (Zero) re-zeros the Basic and Gross readout displays. The

function of this key can be changed by setting the F2 function parameter. For
data entry, this key is also used to decrement a parameter value.

Key

The

key (Print) causes a Print Ticket output. Prints occur regardless of

scale motion. The function of this key can be changed by setting the RST
function parameter. For data entry, this key is also used to scroll a parameter
value by x1000 factor when held with the Tare and Zero keys.

DISPLAYS

BASIC WEIGHT DISPLAY

The Basic Weight Display provides a direct indicator readout. The display

occurs when the Tare Weight =0. 

BRUTTO/GROSS WEIGHT DISPLAY

This Display indicates the Brutto/Gross weight. It is only possible to select

this display mode when Tare > 0, otherwise it is locked out by the indicator.
“B/G” illuminates for this display mode.

NET WEIGHT DISPLAY

This Display indicates the Net weight. It is only possible to select this display

mode when Tare > 0, otherwise it is locked out by the indicator. If after the Tare
Weight is set > 0, the Net Display is selected by the indicator. “NET” illuminates
for this display mode.

TARE WEIGHT DISPLAY

This Display indicates the Tare weight. When this display is selected, it is

possible to directly key-in the Tare Weight value. “T” illuminates for this display
mode. It flashes during tare weight key-in entry.

!"!"!"!"!

Summary of Contents for MP30

Page 1: ...uts implemented on Plug in cards The Plug in cards provide dual FORM C relays 5 A quad FORM A relays 3 A or either sinking or sourcing quad open collector logic outputs The setpoint alarms can be configured in modes to suit a variety of control and alarm requirements High and low absolute high and low deviation and band acting Balanced or unbalanced hysteresis On and off delay timers Auto reset or...

Page 2: ...RS485 Isolation To Sensor User Input Commons 500 Vrms for 1 min Working Voltage 50 V Not Isolated from all other commons Data 7 8 bits Baud 300 to 19200 Parity no odd or even Bus Address selectable 0 to 99 Max 32 meters per line RS485 Transmit Delay Selectable for 2 to 50 msec or 50 to 100 msec RS485 17 ANALOG OUTPUT Types 0 to 20 mA 4 to 20 mA or 0 to 10 VDC Isolation To Sensor User Input Commons...

Page 3: ... parameter and indexes to the next During Digital Tare this key stores the tare weight value keyed in KEY The Tare In Tare Out key both acquires and clears the Tare weight value The function of this key can be changed by setting the F1 function parameter For data entry this key is also used to increment a parameter value KEY The key Zero re zeros the Basic and Gross readout displays The function o...

Page 4: ...as a basic millivoltmeter The basic indicator readout can then be post scaled to read out in weight units The indicator provides two ways in which to scale the display Key in Key in the input and display scaling points using known data Apply Apply the actual input value and key in the corresponding display value The indicator records the input value applied For processes that require linearity com...

Page 5: ...ons Each time the batch function is executed the weight is added to the total and then zeroed PARAMETER LOCK MODE A user input can be used to lock the parameter list When the user input is active the indicator is in the protected parameter mode where it is only possible to access the setpoint values and the security code It is possible to lock the parameter list without using a user input as a pro...

Page 6: ...deviation or band acting the effective trigger point is offset by SP1 In these modes SP1 acts as a master setpoint that the other setpoints track An alternate list of setpoint values can be stored and recalled as needed Select the alternate list with a function key or user input and program the setpoint values This allows for quick changeover for a new process condition Setpoint Alarm Hysteresis a...

Page 7: ...eset Active Reset LIst Select main or alternate setpoint list r 1 Reset setpoint 1 r 2 Reset setpoint 2 USr 2 USr 1 User Input Function no no function pLOC Lock parameter list USr 3 rEL Acquire zero Settings 0 3 8 15 Active Reset Active Function Active Min Active Function Active Max Active Enable Active Enable Active Reset Active Total Active edge Batch Release from active Synchronize Active Hold ...

Page 8: ...ale Factor sCfaC DESCRIPTION COMMENTS RANGE AND UNITS PARAMETER DISPLAY Tracking on reading band Tracking off reading 0 reading band PARAMETER MODULE 6 Setpoint Parameters DISPLAY PARAMETER RANGE AND UNITS DESCRIPTION COMMENTS act n Action for Setpoint ab HI absolute high balanced hysteresis SPSEL Select Setpoint Select setpoint for configuration Select operating action for setpoint See Figures fo...

Page 9: ...sIN Output Source Assignment HI Max reading Select source of linear output signal If totalizer scaling limited to lower 5 digits tot Totalizer value 0 20 0 to 20 mA NEt Net weight b 6 Brutto Gross weight LO Min reading YES Enter sub menu t Tons NO None NO Disables parameter 0 False zero 6 Grams Print Options Line Print Units Suffix Opt UNIt tot Totalizer values 0oc Zero and ounces 6roSS Brutto Gro...

Page 10: ...ers 5 tOt Totalizer Parameters 8 Out Analog Output Parameters 7 SrL Serial Communication Parameters DISPLAY PARAMETER USER SETTING FACTORY SETTING B 6 tArE BRUTTO GROSS DISPLAY TARE DISPLAY rEd Ent SP 3 SP 1 SETPOINT 3 SETPOINT 1 LOC LOC SP 4 SP 2 SETPOINT 4 SETPOINT 2 LOC LOC CodE SECURITY CODE 00000 OPEr HI t MOPERATING MODE MAX CAPTURE DELAY TIME b 6Nt 0000 0 LO t MIN CAPTURE DELAY TIME 0000 0 ...

Page 11: ...rivers pens pencils etc to operate the keypad of the unit EMC GUIDELINES The indicator has been designed and tested to meet the requirements of EMC Directive 89 336 EEC However successful installations depend upon several factors These include cable routing termination of cable shields associated equipment etc Installation The indicator meets NEMA 4X IP65 requirements for indoor use when properly ...

Page 12: ...med as a low acting alarm triggering on the Net liquid weight Each filling operation is started with a Tare Out Tare In sequence This opens the filling valve and when the target setpoint is reached the valve is shut off The valve remains off until a tare operation is repeated thereby restarting the filling operation The fourth output is connected in series with the fill output to disable filling d...