7.2 Procedure of servo calibration
7.2.1
Servo calibration low
Weighing calibration
see Chapter
Step 1 Register conigurati-
on
see Chapter
Rated motor rotating speed, maximum input pulse
frequency, command pulse method, command pulse
polarity, command pulse logic, input pulse per rota-
tion and output pulse per one rotation are registe-
red.
Step 2 Input efective state
see Chapter
Set the efective state of alarm input signal, original
point sensor, ready, limit of OPEN side and limit of
CLOSE side.
Step 3 Manual speed cali-
bration
see Chapter
Calibrate the gate speed for high speed and slow
speed.
Step 4 Positional calibration
see Chapter
The original point and the inal point are calibrated
after selecting the detecting method of the original
point from TYPE1, TYPE2 and TYPE3, and the inal
point from TYPE1 and TYPE2.
Step 5 Manual operation
see Chapter
Manual gate operation is executed.
Step 6 Software limit
see Chapter
Set the software limit of the OPEN side, CLOSE side
and delection.
7.3 Registration of equipment coniguration (CONFIGURATION)
The speciication value and the set value of the servo ampliier and the servo motor used
are registered.
7.3.1
Switch to registration screen of equipment coniguration
Switch from weighing screen to registration screen of equipment coniguration by the
following operations.
7.3.2
Rated motor rotating speed (RATED MOTOR SPEED)
Set the rated motor rotating speed [min
-1
] of servo motor used.
The setting range is 500 to 10000 [min
-1
].
3000 [min
-1
] is set as default.
1. Select [MENU].
2. Select [SERVO CALIBRATION] at menu screen 1/2.
3. Select [CONFIGURATION] at servo calibration screen.
1. Select [RATED MOTOR SPEED] at coniguration screen.
Weight Controller CSD-918
7 Servo calibration
EN-54