UA-66 Urine Analyzer Operator’s Manual
- 12 -
3 Setup
At the main menu( See Figure 2.5 ), use
▲
and
▼
as necessary to select “Options” , and
press
START
to enter “Options” menu:
Options
Strip
Printer
Critical Value
Serial Port
Grade
Serial No.
Realtime Clock
Figure 3.1
Using this menu, the operator can:
♦
Correct the system clock;
♦
Set the options for printer and serial port;
♦
Select the reagent strip , testing results output unit and critical values;
♦
Show serial No. of the instrument.
3.1 Reagent Strip
NOTE: The name of the reagent strip being used must be the same as the reagent strip
selected by the “Options>Strip” menu. Use of any other strip will cause erroneous results.
At the “Options” menu( See Figure 3.1 ), use
▲
and
▼
to select “Strip”, press
START
to enter
the “Options>Strip” menu:
Options>Strip
10 Para.
11 Para.
Figure 3.2
Use
▲
and
▼
as necessary to select the designated reagent strip, press START to validate the
new selection.
3.2 Critical Value
Critical values are a set of values to be compared with the testing results. If the result of a
certain parameter is the same as or higher than the corresponding critical value, a “*” will be
inserted in front of the corresponding row while the report being displayed and printed.
At the “Options” menu, use
▲
and
▼
to select “Critical Value”, press
START
to enter the
Summary of Contents for UA-66
Page 1: ...UA 66 Urine Analyzer Operator s Manual ...
Page 2: ......
Page 6: ...UA 66 Urine Analyzer Operator s Manual iv ...
Page 8: ......
Page 12: ...UA 66 Urine Analyzer Operator s Manual 4 ...
Page 25: ...UA 66 Urine Analyzer Operator s Manual 17 ...
Page 29: ...UA 66 Urine Analyzer Operator s Manual 21 ...
Page 35: ...UA 66 Urine Analyzer Operator s Manual 27 ...
Page 39: ...UA 66 Urine Analyzer Operator s Manual 31 ...
Page 40: ...P N 0103 20 61103 4 0 ...