Disassembly
Repair and Disassembly
4 - 20
4.3.12
Disassembling the NIBP Module
4.3.12.1
Removing the NIBP Pump
1.
Unplug the pump cable connector from the NIBP board.
2.
Disconnect the tubing.
3.
Cut the two tie wraps holding the NIBP pump as shown in FIGURE 4-35.
4.
Remove the pump.
FIGURE 4-35
Cut two (2) tie wraps.
4.3.12.2
Removing the NIBP Dump Valve
1.
Unplug the dump valve cable connector from the NIBP board.
2.
Disconnect the tubing.
3.
Remove the two (2) screws on the bottom of the NIBP rack as shown in FIGURE 4-36.
4.
Remove the dump valve.
FIGURE 4-36
Remove two (2) screws.
Tie Wraps
Pump Cable Connector
Pump Tubing
Screw
Screw