16
12
Image Optimization: Power
1. Select a high-quality image during B mode or B+ Color scanning, and adjust to
place the area of interest in the center of the image.
2. Press
to enter B+Power mode. Use the trackball and
to change
position and size of the Region of Interest (ROI).
3. Adjust the image parameters to obtain optimized images.
4. Perform other operations (e.g. measurement and calculation) if necessary.
Control
Panel
Gain
Changes the Power mode gain---- rotate <iTouch>
knob.
Invert
Inverts the color bar so as to change the flow display
manner.
Menu
Dynamic Range
Adjusts the transformation of echo intensity into color
signal.
B/C Wide
Sets and constrains the maximum width of the B mode
image to that of the Power ROI.
Dual Live
Turns on/ off a synchronous display of B image and
Power image.
B Display
To turn on or off the B image display as Power imaging
remains active.
Frequency
Selects the frequency value in doppler mode.
Scale
Changes PRF to adjust the velocity display range.
A. power
Selects the acoustic power value.
Priority
Sets the flow priority to play: grayscale or color signal.
Packet Size
Adjusts the sensitivity of detecting low-velocity flows.
Power IP
Selects among groups of image combination
parameters to optimize the image.
Map
Selects among groups of image display effects,
including Power maps and Directional Power maps.
Wall Filter
Filters out signals of low speed flow.
Line Density
The function determines the quality and information of
the image.
Smooth
Rejects noise and makes the image smooth.
Persistence
Smooth processing of an image.
Focus Position
To adjust the Power focus position against the ROI
position.