4-7
4.2.13 Complete
If [
VTBI
] is not set during the infusion, when the time in which the remaining liquid
needs to reach [
Time Near End
],
the [
Syringe Near Empty
]
alarm will be triggered,
and this alarm can be cancelled automatically until the syringe is empty. Set [
Time
Near End
], please refer to
6.6 Time Near End
.
When the remaining infusion time is close to the [
Time Near End
]
set by the users,
the [
Time Near End
] alarm will be triggered. If no action has been taken, the alarm
will not be cancelled automatically until the infusion is completed, and then switch to
[
VTBI Complete
] alarm. After KVO mode runs for 30 minutes, the [
KVO Finish
]
alarm will be triggered. Set KVO rate, please refer to
6.1 KVO
.
4.2.14 Standby
Under non-running status, press
to enter the standby countdown shutdown
interface. Press
to modify standby time (range is 00:01-99:59 h:min). The
pump cannot be put in standby mode if there is a high-level alarm.
When the standby state is ended, the title bar will display [
Standby Time Expired
],
press
or
to cancel alarm.
4.2.15 Turn off the Pump
Follow the steps below to turn off the syringe pump:
1. Disconnect from the animal;
2. Hold down
, until the Turn Off progress bar complete, and the power will
turn off.
NOTE
When powering off normally, the current operating data and saved data
will be autosaved.