4-3
Load syringe according to the following method:
1. Open the syringe fixation clamp,
squeeze the handle to open the
clip, and then move the slider to
the appropriate position.
2. Align the syringe flange with the
slot, then load it into the slot, and
clamp the syringe.
(Tip: Flange location)
3. Align the slider to and snugly
against the syringe plunger,
making the clip clamps firm to the
thumb rest. Close the syringe
clamp gently.
WARNING
The flange of the syringe should be firmly loaded into the slot, and not
jutting on the outside of the flange plate.
Before using this syringe pump, the syringe pump, syringe and other
accessories should be installed correctly.
Before using this syringe pump, the brand and specifications of the
syringe used must be confirmed.The brand of syringe pump should be
calibrated on the equipment. If there are no settings for the syringe
used, the rate and the alarms may not be accurate.
4.2.3 Change the Syringe
Follow the steps below to change the syringe:
1. To prevent animal injury, before changing the syringe during infusion, press
to stop the pump.
2. Open the syringe fixation clamp, squeeze the handle to open the clip, move the
slider to the appropriate position, and take out the loaded syringe.
3. Please refer to
4.2.2 Load the Syringe
to reload the syringe.