ELECTROMAGNETIC COMPATIBILITY
GUIDELINES AND MANUFACTURER‘S DECLARATION – ELECTROMAGNETIC IMMUNITY
NOTE 1 With 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not be applicable in all cases. Electromagnetic radiation is affected through
absorption and refl ection of structures, objects and people.
a
The fi eld strengths of stationary transmitters, such as, e.g. the base station of wireless telephones mobile
radios, amateur radio stations, AM and FM radios and TV transmitters cannot theoretically be accurately
predetermined. In order to determine the ELECTROMAGNETIC ENVIRONMENT with regard to the stationary
transmitters, a survey of the electromagnetic phenomena on site should be considered. If the measured fi eld
strengths at the site used on
MIMLED 600
and
MIMLED 1000
exceeds the above-mentioned
COMPLIANCE LEVEL,
MIMLED 600
and MINOR SUR-GICAL LIGHTS 100
000 LX should be observed to verify
intended function. If unusual performance charac-teristics are observed, additional measures may be
required, such as, e.g. a modifi ed alignment or another location of
MIMLED 600
and
MIMLED 1000
.
b
The fi eld strengths should be less than 3V/m over the frequency range of 150kHz to 80MHz.
11.3 RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND
MOBILE RF TELECOMMUNICATION EQUIPMENT AND THE DEVICE
(NOT LIFE-SUPPORTING)
THE RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF TELE-
COMMUNICATION DEVICES AND
MIMLED 600
AS WELL AS
MIMLED 1000
MIMLED 600
/
MIMLED 1000
is intended for use in an electro-magnetic environment where radiated
RF interferences are controlled. The customer or the user of
MIMLED 600
/
MIMLED 1000
can help to
avoid electromagnetic distur-bances by maintaining a minimum distance between portable and mobile RF
telecommunication equipment (transmitters) and
MIMLED 600
/
MIMLED 1000
– as specifi ed below
according to the output power of the telecommunication equipment
RATED OUTPUT
OF THE
TRANSMITTER (W)
SEPARATION DISTANCE ACCORDING TO THE TRANSMITTER FREQ.
(M)
150 kHz to 80 MHz
d = 1.17√P
80 MHz to 800 MHz
d = 1.17√P
800 MHz to 2.5 GHz
d = 2.34√P
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.69
3.69
7.38
100
11.67
11.67
23.33
For transmitters rated at a maximum output not given in the above table, the recommended separation dis-
tance d in meters (m) can be estimated using the equation in the respective column where P is the maximum
rated output of the transmitter in watts (W) according to the manufacturer‘s specifi cations.
NOTE 1: With 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not be applicable in all cases. Electromagnetic radiation is affected through
absorption and refl ection of structures, objects and people.