XRS-5 Transducer
PL-590
Applications
Transducer Placement
The following graphic shows the best placement of the XRS-5 transducer.
Locations
A.
Primary location.
This is the preferred location and should be used
whenever possible. The centre of the tank generally gives the most reliable
readings because there are fewer obstructions to provide false echos.
B.
Alternate location.
This location is used if the centre of the tank is already
in use or if the tank roof is too weak to hold the transducer safely.
C.
Poor location.
This is a poor installation location. The echos are shown
reflecting away from the transducer face.
beam
angle
rise
‘B’
‘A’
‘C’
may require target to
obtain empty reading
discharge
tank
manufacturer’s
empty level
Empty level for ‘B’
location. Below this
level, echo would
reflect away from
the transducer.
Maintain full fluid
level for full or
offset calibration.
Do not allow
material to enter
blanking zone.
span:
distance between
Empty and Full
levels in the
measured process
run
Summary of Contents for PL-590
Page 1: ...Instruction Manual PL 590 XRS 5 Transducer ...
Page 4: ...Page 4 XRS 5 Transducer PL 590 ...
Page 6: ...Page 6 XRS 5 Transducer PL 590 Specifications ...
Page 10: ...Page 10 XRS 5 Transducer PL 590 Installation ...
Page 18: ...Page 18 XRS 5 Transducer PL 590 Applications ...
Page 21: ...PL 590 XRS 5 Transducer Page 21 Installation Diagram Installation Diagram ...
Page 22: ...Page 22 XRS 5 Transducer PL 590 Installation Diagram ...