Elix® Essential 3, 5, 10, 15 (UV) / RiOs™ Essential 5, 8, 16, 24
39
Basic troubleshooting
Icon or error
What to do?
•
Product water resistivity < set point or RO permeate water
conductivity > set point.
•
Verify the consumables have been replaced.
•
Low feed water pressure is detected (0.5 bar).
•
Verify the feed water faucet is opened.
•
Verify the feed water pressure is in specification.
•
Verify the feed water pressure regulator is well adjusted or is working
properly (if installed).
•
Water leak is detected.
Er:02
•
Feed water temperature > 35°C.
•
Verify that the feed water temperature is really greater than 35 °C. See
the feed water quality requirements to operate the system in compliance
with the manufacturer specifications.
Er:20
•
Network cable is unplugged: cancel the warning or plug the Ethernet
cable.
Er:24
•
Replace UV lamp 254 nm: the UV lamp is exhausted.
•
Contact your Service Representative for replacing the UV Lamp 254 nm.
Er:26
•
Replace UV lamp ASM: the UV lamp is exhausted.
•
Contact your Service Representative for replacing the UV Lamp ASM.
Summary of Contents for Elix Essential 10
Page 1: ...User Manual Elix Essential 3 5 10 15 UV RiOs Essential 5 8 16 24...
Page 2: ......
Page 8: ......
Page 9: ...Table of contents...
Page 12: ......
Page 13: ...Items needed...
Page 16: ......
Page 17: ...System...
Page 26: ......
Page 27: ...Consumables replacement...
Page 30: ......
Page 31: ...Sanitization and cleaning...
Page 34: ......
Page 35: ...Other functions...
Page 38: ......
Page 39: ...Error messages...
Page 42: ......
Page 43: ...Chemical and technical specifications...
Page 46: ......
Page 47: ...Legal information...
Page 50: ......
Page 51: ...Item catalog numbers...
Page 54: ...Elix Essential 3 5 10 15 UV RiOs Essential 5 8 16 24 FTPF12381 V2 0 04 2012...