![Millipore DDHW00010-WW User Manual Download Page 48](http://html1.mh-extra.com/html/millipore/ddhw00010-ww/ddhw00010-ww_user-manual_1796888048.webp)
Direct Detect® Spectrometer User Guide
44
14.6 Direct Detect® System IQ/OQ/PQ Report Summary
Reference
Section
Checklist
Result
Pass
Fail
14.1
Verify power and cable connections
● Power adapters are connected to Netbook computer and
Direct Detect® instrument
● Verify connection between Netbook computer and Direct
Detect® instrument (cross over cables)
Verify Software communication
● Direct Detect® Software launches successfully
● Login successful
● Instrument detected
14.2
Instrument Hardware Diagnostic
● All functions have green check mark
14.3
System OQ
● OQ test report indicates
Overall Test Result = PASSED
14.4
System PQ
● PQ test report indicates
Overall Test Result = PASSED
14.5
Verification of Protein Measurement
● Direct Detect® measurement and UV reader measurement
are within 10% of one another for a 2 mg/mL BSA standard.
Testing completed by:
Date: