Direct Detect® Spectrometer User Guide
10
Type
Definition
6H CFG switch
CFG is a four-fold DIP switch:
#1 ON and #2 OFF: DHCP mode - use last DHCP address
#1 OFF and #2 ON: use ON-IP address 10.10.0.1
#3: stay in boot loader, do not load the application program
#4: reserved
The abbreviation CFG means
Configuration
.
6I SBY/RES button Pressing this button for approximately 2 seconds, activates/
deactivates the stand-by mode. Pressing this button for more
than 6 seconds resets the spectrometer to the original state. The
abbreviation SBY means
Standby
and RES means
Reset
. Do not use
unless instructed to do so by service personnel.
6J POWER port
This port connects the spectrometer to the power supply via the
supplied power adapter and power cord. The POWER port is for
extra low-voltage connectors. Use only power cords/supplies that
have been supplied by EMD Millipore Corporation.
6K Kensington lock
slot
The lock slot is an anti-theft lock for mobile hardware. Lock and
cable are not supplied.
7.2 Direct Detect® Computer and Software
The Direct Detect® spectrometer is designed to operate with the Direct Detect® Software
that is pre-installed on the Netbook computer. For computer specifications, refer to
section 15.
Direct Detect® Software may be updated periodically. For information on the most up-to-
date software, go to www.millipore.com/directdetect. Register your instrument in order to
receive notification about relevant software upgrades.
WARNING:
Do not install third-party software (screen savers, etc.), as this may impact the
performance of the Direct Detect® system.
7.1 Direct Detect® Spectrometer General Technical Features
,
continued