Turn the control knob clockwise to increase the speed.
Turn the control knob counterclockwise to reduce the speed.
Using the Timer in Standard Mode
To adjust the value of the timer:
1.
Press and hold the control knob until the value of the timer is blinking.
NOTE
The timer value cannot be adjusted when the pump is running.
When using the timer, only the pressure monitoring mode is available. If
the pressure regulation mode is activated at the same time the timer is
activated, a message displays: “Using the Timer will switch the Pressure
Measurement System from Regulation to Monitoring.” Turn and press the
control knob to select
Continue
.
2.
Turn the control knob clockwise to increase the timer value.
Turn the control knob counterclockwise to decrease the timer value.
Press the control knob to confirm the value.
NOTE
The timer value varies from off (deactivated) to 999.
Between 0.5 and 9.9, the display is 1 digit and 1 decimal point. Between 10
and 999, the display is 2 or 3 digits without a decimal point.
3.
To start the pump, press the control knob or the footswitch (if connected).
The pump starts. A progress bar shows the remaining time in the
Timer
area. When the progress bar is full blue, the cycle is complete, and the pump
stops.
Timer value
Progression of the cycle