Page 49 of 113
Information Area
This area is used for the display of different messages:
•
Possible actions using
and
buttons
•
Information and instructions regarding test method step (in Test Method Mode)
•
Alert messages and error messages
Filtration Parameters
Timer
The timer function stops the pump automatically after a preset filtration time.
The timer value can vary from off (not used) to 999 (see
).
NOTE
The timer default value is
off
.
Speed
The pump speed value can vary from 1 to 150.
NOTE
The default speed value is 75.
Pressure
The Steritest
®
Symbio Pump constantly monitors the pressure inside each
Steritest
®
canister and indicates when pressure goes above normal working
conditions. Colored bars and audible beeps indicate the pressure level. A white
bar indicates that the pump head is not running.