background image

Milli-Q® HX 7040-7150 / HR 7060-7220

User Manual

Summary of Contents for HR 7060

Page 1: ...Milli Q HX 7040 7150 HR 7060 7220 User Manual...

Page 2: ......

Page 3: ...iew and Workspace Screens 18 Overview Screen 19 How to Change the System from Standby to Ready 23 Workspace Screens 25 Glance Workspace 26 Maintenance Workspace 30 Configuration Workspace 32 How to Vi...

Page 4: ...0 Consumable Catalogue Numbers 51 Accessory Catalogue Numbers 52 System Catalogue Numbers 53 Appendix 54 Display Icon Description 55 System Operating modes 58 Emergency Procedure in Multi system confi...

Page 5: ...arrant these systems for any specific application It is up to the end user to determine if the quality of the water produced by our systems matches his expectations fits with norms legal requirements...

Page 6: ...lectrical ground connection is made This ELECTRICAL HAZARD sticker is used to refer to a position on the system cabinet or inside where an electrical danger could exist This HOT SURFACE sticker is use...

Page 7: ...ed electric or electronic equipment Proper disposal of equipment containing electric or electronic components will help to reduce pollution effects to the environment or to human health Proper recycli...

Page 8: ...7150 HR 7060 7220 unless specified otherwise in the text The abbreviations LC and HC used after HX 7040 7080 and HR 7060 7120 refer to systems for Low Chlorine and High Chlorine feed water The words d...

Page 9: ...Specifications Specifications 9...

Page 10: ...y Index SDI 7 Total Organic Carbon TOC 1 ppm Free chlorine for LC systems 1 5 ppm Free chlorine for HC systems 1 5 ppm 3 ppm 12 when the optional UF pretreatment unit is installed Product Water Specif...

Page 11: ...Analytical laboratory use GB T6682 2008 Grade 2 Water for Analytical laboratory use ASTM D1193 type 2 2006 Reapproved 2011 Reagent Grade Water Japanese Industrial Standard JIS K 0557 A3 2008 European...

Page 12: ...System Dimensions HX and HR System dimensions 12 Specifications...

Page 13: ...e the consumable or any accessories The UF pre filtration option weight is 22 kg 121 3 lb Electrical Specifications The systems can be supplied by Mains electricity 90 253 VAC with a frequency range f...

Page 14: ...ude 2000 m Ambient operating temperature 10 40 C Relative humidity 80 up to 31 C decreasing linearly to 50 relative humidity at 40 C Ambient storage temperature 0 40 C Ambient storage humidity 10 95 I...

Page 15: ...on label 5 Progard pack s 16 Inlet connector 3 4 BSP 6 Wheel locking nut 17 Outlet connector 3 4 Tri Clamp 7 Securing bracket 18 External pretreatment port 8 Water system backpanel 19 Leak detector po...

Page 16: ...III HX 7040 flow schematic HR 7060 flow schematic This view and more information about components can be retrieved using the Flow Schematic application at Glance Workspace Flow Schematic The volume o...

Page 17: ...system is connected to the external sensors and actuators linked to the distribution loop The pump in simplex or duplex mode Tank level mangement The ASM UV lamp Dumping valve Leak detection Overflow...

Page 18: ...on contain the system applications How to Navigate Between Overview and Workspace Screens Procedure Navigation between the Overview screen and the Workspace screens is done using the display icon butt...

Page 19: ...tribution process is in RECIRCULATION mode The Distribution process can also be NOT ACTIVE or in MULTI SYSTEM auxiliary mode The system operating modes are described in the appendix C Measure Widgets...

Page 20: ...he water from the distribution loop for example after a loop sanitization or prior a sampling procedure use Distribution in Scheduled mode The Distribution is activated based on defined schedule Note...

Page 21: ...installed Without distribution kits installed With distribution kits installed To access the distribution modes click on distribution mode button The distribution panels will show the three distribut...

Page 22: ...gards can be displayed D2 Vent filter status and gauge D3 Loop filter status and gauge E Notification and Navigation bar E2 Notification button showing date and time Colour changes depending on system...

Page 23: ...EADY mode the operating mode will display STANDBY 1 The Makeup and Distribution processes can be changed between STANDBY and READY using the three horizontal bars at the top of their respective window...

Page 24: ...nding process enters PLEASE WAIT Results When both Makeup and Distribution processes are in Ready the filling of the tank and loop management is automatic For more information please refer to System O...

Page 25: ...Workspace Screens There are three Workspace screens Glance Maintenance and Configuration which list all available applications System Display 25...

Page 26: ...everal electronic boards using different software You may need to provide some or all of this information to your Service representative in case of troubleshooting Software and firmware board versions...

Page 27: ...e On HX systems the Elix product conductivity and temperature are monitored Distribution stage If the kits are installed TOC Distribution Resistivity and temperature are displayed Hydraulic measuremen...

Page 28: ...eplacement date are displayed UV lamps Catalog number installation and replacement date are displayed for the following components Makeup on HX systems only and if installed Automatic Sanitization Mod...

Page 29: ...the system manages a distribution loop Auto recirculation distribution loop automatically recirculates once per hour during the period scheduled Auto recirculation cycle duration can be adjusted from...

Page 30: ...same consumable status information in the Glance Workspace Consumables Application but you cannot perform consumable replacement from there Consumables found in this application Progard pack s and if...

Page 31: ...opening this application two dates related to cleanings are shown The Performed date indicates the last time a cleaning was done The Due date indicates when the next cleaning is due Note The cleaning...

Page 32: ...ilar information can be seen but not changed Information Application allows modification of system information Product information Parameters from the manufacturer that uniquely identify the system ca...

Page 33: ...or can be activated here Measure outputs Two measures can be selected from the list of parameters these parameters can be monitored by external devices Units Set your desired units here Alarm and Aler...

Page 34: ...selected Export Import Application The system configuration file can be exported from this application It is highly recommended to export the system configuration file and keep it as a backup in order...

Page 35: ...ge to access to the other available applications On the left and right side of the application windows two arrows are present Click on these arrows to navigate through different application windows Cl...

Page 36: ...nues to operate Replace Progard in xx days Alarm An alarm is displayed in red meaning that a problem was detected by the system The system continues to operate Elix Resistivity Set point Alarm Stop An...

Page 37: ...t be cleared unless the cause of the alarm has been fixed Acknowledging Alerts clears the message for 24 hours Procedure To acknowledge an alarm or alert 1 Tap on the Alarms and alerts Notification bu...

Page 38: ...and the screensaver displays the main system information Screensaver visual states STANDBY Makeup process standby and or Distribution process are in STANDBY READY Makeup process and Distribution proce...

Page 39: ...Maintenance Alerts are triggered when a Consumable Replacement Cleaning or Sanitization is required Maintenance 39...

Page 40: ...e launched using the Consumables application or the Sanitization Cleaning application See Applications at the Maintenance Workspace chapter 1 Tap on the Alerts Notification button on the Overview scre...

Page 41: ...l and one RO Regeneration Plug Tool Two Chlorine tablets RO Protect C 2 Alternatively if you can use Progard Auto clean Packs that already contains Cl2 cleaning agent What do you need to perform an RO...

Page 42: ...due to mineral scaling The use of RO Acid Cleaner removes some or most mineral scale buildup on the surface of the RO membranes RO Base Cleaner RO Care B Delivered in a pouch The empty pouch is remove...

Page 43: ...as an example You can access to the same applications and screen views on the remote display interface and the system s main display Note When three users are connected the system informs any new con...

Page 44: ...or Network configuration use the System Settings Application About this task Important The network configuration should only be changed from the system display Do not attempt to change the settings re...

Page 45: ...If you use DHCP mode check the DHCP box If using static mode you must complete the three fields IP address Subnet mask and Default gateway using ip V4 format 5 Validate your configuration by pressing...

Page 46: ...Power Outlet Note Do not power off the system by removing the power cord from the Power Outlet If you need to power off the water system for any reason 1 Put the Makeup and Distribution processes into...

Page 47: ...Distribution Schedule application 2 The Distribution Schedule application opens the bullet points in red will guide you throught the following steps a Select one day of the week from the right side s...

Page 48: ...g Once a day has been completed a profile day can be copied to another day of the week using the copy function 5 48 How to Use the Distribution Schedule...

Page 49: ...makeup process will be automatically stopped by the application 1 From the Maintenance workspace launch the Sanitization Cleaning application and select Tank emptying 1 2 Launch the tank emptying func...

Page 50: ...Ordering Information 50 Ordering Information...

Page 51: ...abel Catalogue Number Description RO Protect C ZWCL01F50 CL2 tablets Qty 50 RO Protect C USA only 5874316024 CL2 tablets Qty 24 RO Protect C Canada only 5874316024C CL2 tablets Qty 24 RO Care A ZWACID...

Page 52: ...xternal Pressure Regulator ZLXL000PR Feed water pressure regulator 0 25 bars UF Pretreatment Unit ZUFPREUN0 Ultrafiltration filter Unit 3 4 UF Installation kit ZUFPREUN8 Installation kit for the UF Pr...

Page 53: ...ter Purification System 120V 60 Hz ZLXL52150 Milli Q HX 7150 Water Purification System 230V 50 60 Hz ZR0L72060 Milli Q HR 7060 Water Purification System HC 100V 50 60 Hz ZR0L62060 Milli Q HR 7060 Wate...

Page 54: ...Appendix 54 Appendix...

Page 55: ...selected item s from a list Modifies the selected item s of a list Opens the Glance Workspace Opens the Maintenance Workspace Opens the Configuration Workspace Shows water quality measurements Shows h...

Page 56: ...icons to indicate the connection status via Ethernet or the front side USB port As an example for the Overview screen USB port front side No USB device detected on the system A USB device has been ins...

Page 57: ...Ethernet connection ask the user having the remote IP address to leave the Maintenance and or Configuration applications A user is using the MMI directly and is in the Maintenance and or Configuratio...

Page 58: ...To clean RO Membrane s REAGENT REMOVAL To remove cleaning agents after CL2 CLEANING or pH CLEANING ALARM STOP Stops the Makeup process in the event of an Alarm Stop signal BACKWASH FILTER REGENERATIO...

Page 59: ...get your primary system up and running 1 Power off the primary system using the main switch on the system backpanel 2 From the primary system backpanel disconnect the three external connectors C1 C2 O...

Reviews: