ULTRASONIC PORTABLE MILK ANALYZER
www.milkotester.com
13
display. If the unit is connected to the printer all results will be
automatically printed out.
At this point the sample cup can be removed from under the pipette.
The instrument will continue to show the measured values until a new
analysis is initiated.
WARNING!
Make sure that the instrument is at rest during analysis. Any shaking of
the device will cause inaccurate results.
NOTE:
Due to the possible presence of water inside the flow system left from
the last flushing procedure, we recommend that you do not take into
account the first analysis after flushing if they appear inaccurate. To
prevent or reduce this inaccuracy due to water in the system, please
refer to chapter “Cleaning and Maintenance”, section “Thorough
Cleaning”.
To access the menu, press the MENU button. The range of functions
is grouped into submenus. Scroll through the menu to select the one
that you need and then select the settings you need to make. Press
ENTER to activate them. Press EXIT to keep the previous settings.
4.3 MAKING CORRECTIONS AND CALIBRATION
In the process of work with the analyser there is a possibility the
results to start differing between the data for some of the measuring
parameters when measured with the milk analyzer and the
corresponding reference method of analyses (Gerber for fat, Kjeldhal
for proteins etc). In order to establish the possible discrepancy and to
correct the readings of the milk analyser do the following:
Taking samples and preparation of samples for checking the accurracy
of the milk analyser, making corrections and recalibration
This is a basic moment for the correct checking the accuracy of the
analyser and for making correct and precise correction and calibration.
It is accomplished according Appendix Taking and preparation of
samples for checking correctness of the milkanalyser, making
corrections and recalibration.