Programme settings
99
Adjusting programme settings
The programme settings should be adjusted to suit technical
requirements and the load.
Additional specialist knowledge is required to alter programme
settings and this should therefore be undertaken only by experienced
users or by the Miele Customer Service Department.
Programme and dispensing changes must be documented for
validated processes. In Germany this is a requirement of the
Medical Devices Operator ordinance (MPBetreibV).
The process must be validated again as necessary.
Programme structure
Each programme is subdivided into programme blocks which run one
after another. A programme consists of at least one and a maximum
of eleven programme blocks. Each block can occur only once in a
programme.
The so-called programme header is placed above the programme
blocks and contains general programme settings. Individual wash
block parameters are also globally activated or deactivated here.
Programme
header
-
Change volume of water
Every programme block with water intake is allocated a nominal
water quantity. The water quantity can be raised or lowered
incrementally for all blocks to the base value in the programme
chart.
-
Drain time
If the on-site waste water system is insufficient to drain the waste
water from the wash chamber within the time allocated, the
drainage time can be lengthened by a set amount.
Parameters for measuring wash pressure and spray arm monitoring
can only be accessed by the Miele Customer Service Department.