Programme settings
103
Altering a programme
A programme is altered in two steps:
- Altering the programme begins with a list of all wash blocks
allocated to the programme. First this list must be confirmed.
- Then the programme parameters can be altered.
Use this option to document all changes to factory settings in case
of a subsequent Service call requirement.
Programme and dispensing changes must be documented for
validated processes. In Germany this is a requirement of the
Medical Devices Operator ordinance (MPBetreibV).
The process must be validated again as necessary.
...
Programme settings
Change programme
Change programme
Vario TD Instruments
Vario TD Inst 4trays
Select the programme you want to alter.
For more information see “Allocating wash blocks”.
Allocating wash
blocks
Every programme change begins with a list of the wash blocks.
Vario TD Inst 4trays
Pre-wash 1
Accept
All wash blocks which are allocated to the programme are listed in
the display. The allocation can be adapted as necessary by Miele
Service.
Select the option
Accept
and confirm with
OK
.
Further setting options will then follow. You can edit these in any
order you want.