Replacing filter
plates
Filter plates must be replaced approx. every 2 weeks or after 20
cycles have been run.
Completely unscrew holder AUF 1.
Dismantle the holder and dispose of the filter plate.
Rinse the bottom part of the holder thoroughly before inserting a
new filter plate.
Screw the top and bottom parts of the holder back together and fit
the holder back onto the injector manifold.
Fitting E 499 for suction and rinsing cannulae
The E 499 is for reprocessing rinsing and suction cannulae, including
suction breakers. It comes with its own instructions.