en - Areas of application
42
Remove an injector nozzle from the injector manifold and replace it
with adapter A 801
.
Screw the bottom part of holder
into the adapter.
Insert a filter plate
into the bottom part of the holder.
Do not insert a filter plate in the holder if the injector manifold
has been fitted with an A 800 tubular filter.
Screw the top part of the holder into the bottom part of the holder.
Preparing
transmission
instruments
Before loading check that the silicone adapters are present and
correctly fitted.
Wipe away any deposits from the area around the silicone adapters
that may be present depending on local water conditions using a
damp cloth.
Check that internal sections of transmission instruments are free of
blockages.
Fit the instruments into their holders using the respective silicone
adapters.
After processing dry internal sections of transmission instruments
with sterilised compressed air and then sterilise as required follow-
ing manufacturer's instructions and national health and safety regu-
lations.
Before using transmission instruments again following repro-
cessing, a function check must be carried out, e.g. by spraying into
a basin, to ensure they are clear.