For validation purposes please make sure that you follow the
loading instructions given on the template.
To minimise the risk of injury, the mobile unit should be loaded
from the back forwards and unloaded in the reverse order.
Any hollow items that have become disconnected from their in-
jector nozzles, adapters or holders during processing must be re-
processed.
Before loading the machine and before starting a
programme
– Check that fittings such as irrigation sleeves and injector nozzles
are securely screwed into position.
Make sure that all screw connectors are fitted with injector
nozzles, adapters, irrigation sleeves or blind stoppers to ensure
that all rinsing fittings in use are supplied with sufficient standard-
ised pressure.
Damaged fittings such as injector nozzles, adapters and irrigation
sleeves must not be used.
Rinsing positions not in use do not have to be fitted with blind
stoppers.
– Check that spray arms are rotating freely.
– Check spray arm jets for blockages.
– Check that the basket is correctly connected to the water supply in
the washer-disinfector.