en - Network connection
51
Activating the
configuration
The interface configuration is only effective after the cleaning machine
has been restarted.
Wait approx. 10 seconds before switching the cleaning machine off
for a restart.
Switch the cleaning machine off and then back on again with the
button.
Wait approx. 90 seconds for the module to initialise.
Then check if the interface is working.
Fault 559
If the fault message
Fault 559
is displayed during commissioning or a
programme is interrupted with this message, the RS232 interface is
activated and must be deactivated.
Open the menu for configuring the RS232 interface via
Further set-
tings
/
Interface
.
Select the
Ethernet
option and confirm your choice with
OK
.
The RS232 interface is deactivated. After around 90 s, the module is
ready for operation.