Handles

^

Place contaminated handles on a jet, e.g. E 452.

Veress cannulae

^

Take the Veress cannula apart.

^

Place the outer part of the Veress cannula on jet
E 453 with spring clamp E 472.

^

Screw the inner part of the Veress cannula to a
Luer-Lock adapter

\

, keeping the stopcock open.

Small components

^

Place small components such as seals and
stopcocks in the mesh tray with the led E 907/1.

^

Close the lid, and hang the mesh tray in the mobile
unit.

Trocar sleeves Ø 10 - 20 mm

^

Take the trocar sleeves apart.

Process small components in a mesh tray with a
lid.

^

Attach trocar sleeves with Ø 10 - 15 mm to an
E 454 jet.

^

For trocar sleeves with Ø 15 - 20 mm, use jets
E 464 and E 454 together.

^

Position the trocar sleeve with the spring clamp so
that the water reaches critical areas through the
port in the side of the jet.

^

Connect the trocar tube at the side with hose
adapter E 448.

^

Insert the hose adapter in a silicone adapter.

^

Secure the self-closing valve to an E 454 jet for
trocar sleeves or an E 453 jet with a retaining clip.

^

Position the self-closing valve with the spring clamp
or retaining clip so that the water reaches critical
areas through the port in the side of the jet.

Insert E 905/1 for short MIS instruments