Handles
^
Place contaminated handles on a jet, e.g. E 452.
Veress cannulae
^
Take the Veress cannula apart.
^
Place the outer part of the Veress cannula on jet
E 453 with spring clamp E 472.
^
Screw the inner part of the Veress cannula to a
Luer-Lock adapter
\
, keeping the stopcock open.
Small components
^
Place small components such as seals and
stopcocks in the mesh tray with the led E 907/1.
^
Close the lid, and hang the mesh tray in the mobile
unit.
Trocar sleeves Ø 10 - 20 mm
^
Take the trocar sleeves apart.
Process small components in a mesh tray with a
lid.
^
Attach trocar sleeves with Ø 10 - 15 mm to an
E 454 jet.
^
For trocar sleeves with Ø 15 - 20 mm, use jets
E 464 and E 454 together.
^
Position the trocar sleeve with the spring clamp so
that the water reaches critical areas through the
port in the side of the jet.
^
Connect the trocar tube at the side with hose
adapter E 448.
^
Insert the hose adapter in a silicone adapter.
^
Secure the self-closing valve to an E 454 jet for
trocar sleeves or an E 453 jet with a retaining clip.
^
Position the self-closing valve with the spring clamp
or retaining clip so that the water reaches critical
areas through the port in the side of the jet.
Insert E 905/1 for short MIS instruments