INSTALLATION
REV.0.03_COD.610078_A4
Page 20
Drain cooling
The machine is equipped by a system of drain cooling, which can be useful in order to reduce
the temperature from the standard temperature of 200° to 140°F (93° to 60°C).
3.7
Ambient ventilation requirements
During normal operation, the machine warms up itself dispersing heat and humidity.
Therefore, in order to guarantee a comfortable environment with good temperature and
humidity for the user, it is necessary to prepare an air conditioning or air circulation system
capable to balance the emissions reported in the installation plan.
Only when toxic detergents are used and/or it is expected that it is not possible to guarantee
acceptable levels of temperature and humidity in the installation environment, the air exhaust
of the equipment must be connected to a venting system.
The air that is expelled from the chamber must not be re-circulated so as to protect the user
from any undesired emissions produced by the machine or by the chemical products present
within the work area.
The machine connections are shown in detail on the
installation plan and electrical wiring diagram.