Chapter 1: Introduction
DSS-7000
Test Preparation
Inspecting the Battery
Before starting the test visually inspect the battery for:
•
Cracked, buckled, or leaking case. If you see any of these
defects, replace the battery.
•
Corroded, loose, or damaged cables and connections. Re-
pair or replace them as needed.
•
Corrosion on the battery terminals, and dirt or acid on the
case top. Clean the case and terminals using a wire brush
and a mixture of water and baking soda.
•
Low electrolyte level. If the electrolyte level is too low, add
distilled water to fill up to ½ above the top of the plates
and fully charge the battery. Do not overfill.
•
Corroded or loose battery tray and hold-down fixture.
Tighten or replace as needed.
Testing Out-of-Vehicle
The preferred battery test location is in the vehicle. However, if
you plan to test out of the vehicle:
•
Always disconnect the negative cable from the battery
first and reconnect it last.
•
Always use a carry tool or strap to lift and transport the
battery.
Failure to properly install lead terminal adapters,
or using adapters that are dirty or worn, may
cause false test results.
When testing side-post or Group 31 batteries, always use lead
terminal adapters provided with the tester—do not test at
the battery’s steel bolts. To avoid damage, never use a wrench
to tighten the adapters more than ¼ turn.
Testing In-Vehicle
The preferred test position is at the battery posts. If you must
test at a remote-post location, it should have both a positive
and negative post. Otherwise, you must remove the battery
and perform an out-of-vehicle test.
At the start of the test, make sure all vehicle accessory loads
are off, the key is not in the ignition, and the doors are closed.
Connecting To A Battery
Do not connect the tester to a voltage source
greater than 30 Vdc.
Connect the clamps to the tester: the red clamp to the positive
(+) terminal and the black clamp to the negative (–) terminal.
If you connect the clamps in the wrong polarity (positive to
negative or negative to positive), the tester displays CLAMPS
REVERSED! Reconnect the clamps.
To make sure both sides of the clamps are gripping the
terminals, rock the each clamp back and forth. A poor
connection will prevent testing, and the tester will display
the message CHECK CONNECTION. If the message reappears
after you have correctly reconnected the clamps, clean the
terminals and reconnect.
Connecting An Accessory Cable
If you are using an accessory cable, plug it as you would a
phone jack into the accessories port on top of the tester. It
locks automatically into the port. To remove it after testing,
press the lever and pull the connector out.
Setting User Preferences
Before starting your test you may want to customize the use
of your analyzer by setting preferences in the Settings (
)
Menu. The Settings Menu is described in Chapter 6.
Initial Power Up
1. Select the System, Keyboard, Test Result, Email and Print-
out default language: English, French Canadian, Latin
Spanish. Tap Continue to advance to the next screen.
2. Use the keyboard to create a Username and Password.
Tap Continue to advance to the next screen.
IMPORTANT: By default, the first account to be
established during the initial setup process will
be an Admin account.