MICROTECH
OPERATION INSTRUCTIONS
Clean the oil from the measuring rod and measuring anvil.
Switch on indicator with using
O/ON
button (short press)
Use p.2 of this manual (button functions) to select mode or functions.
Attach indicator in the stand
in the working position in relation
to testing object.
Check the zero setting. Provide tightness
of the measuring rod amount 0.4 mm – 0.6 mm
in relation to testing object. Press button O/ON to zero setting.
Check the stability of the zero position of the 4-5x caging.
Wipe the testing object with a clean soft cloth.
WARNING !
Measuring rod should move without shock at the end of turn.
Avoid contact with the indicator and oil emulsion.
Do not turn indicator when it is fixed in the tool holder by a sleeve
BATTERY REPLACEMENT
If necessary, open the battery cover on the bottom
side of the body, insert the battery (type СR2032)
according to the polarity of the electrodes.
Blinking display information or absence suggests to replace battery.