EN
If you switch the Target ON,
OK
-
+
NEXT
TARGET: ON
0.7
1.5
LO
press the “NEXT” button and the “LO”
message will appear on the display,
indicating the user to set the lower limit of the
target range. Use the “+” and “-” buttons to
change the lower limit. The “LO” message
will also be flashing on your test result display
whenever your result is below the lower limit.
Press the “NEXT” button when you have the correct lower INR limit and
are ready to set the higher limit of the target range.
OK
-
+
NEXT
TARGET: ON
0.7
1.5
HI
The “HI” message will appear on the display.
Use the “+” and “-” buttons to change the
higher limit. The “HI” message will also be
flashing on your test result display whenever
your result is above the higher limit.
Press the “OK” button to accept any setting. The system will save your
setting and will return to the “SETTINGS” screen.
If you switch the Target OFF,
OK
-
+
NEXT
TARGET:
OFF
0.7 1.5
press the “OK” button. The system will
save your settings and will return to the
“SETTINGS” screen. The “LO” or “HI”
message will not be flashing on your test
result.
If you would like to reset the settings, press the “OK” button to go back
to the “SETTINGS” screen. Repeat section 8.2 until the desired setting is
selected.
NOTES:
• In the Setup Mode, pressing the “OK” button will save your settings,
then return to the “SETTINGS” screen.
• If the power is cut off, current settings will be lost and the qLabs
®
ElectroMeter will reset to the factory default settings. To save
settings, install / replace the batteries while keeping the power
adapter connected to the power supply.
14