background image

 
 

 

Compact Numeric Keypad - MP3393

 

 

Page 2 of 9 

 

T_MP3393-Compact  Numeric Keypad Leaflet_V1.3

 

 

Contents of MP3393 Kit 

 

 

1 x MP3415 Compact Numeric Keypad 

 

1 x MP3406 Instructions for Use (this document) 

 

Features and Specifications 

 

 

22 multi-function keys enabling full stimulator control when used in combination with a 
Micropace Touch Screen (MP3114). 

 

High quality mechanical switches with ~20 million keystroke lifetime. 

 

Weight: 235 g (0.51 lb) 

 

Dimensions:  159 x 108 x 38 mm (6.25 x 4.25 x 1.5 in) 

 

Cable length: 2m 

 

CE, FCC, and RoHS approvals. 

 

Introduction 

 

For Stimulators with touch screens, the Compact Numeric Keypad (MP3415) may be substituted for the 
Standard Keyboard to reduce desktop footprint. The user may then use the touch screen or the Keypad 
arrow keys (NumLock OFF) to select pacing protocols, stimulation channels and required extra stimuli, 
and may use the Keypad number keys (NumLock ON) for rapid adjustment or entry of numeric values.  
Starting and stopping stimulation / pacing can be done with the dedicated PACE key. 
 
This Numeric Keypad is intended for use as a permanent replacement to the Standard Keyboard.  It is 
designed, however, to replace the keyboard without any software configuration required.  Therefore, the 
Numeric Keypad and the Standard Keyboard can be swapped as necessary depending on your needs. 
 
Note that the three setup functions of naming protocols, pace sites and user setups require use of the 
alpha keys on the Standard Keyboard. 
 

Compatibility 

1.  Numeric Keypad MP3415 is compatible with EPS320 Stimulator Software 3.21 or above only. 

2.  The Numeric Keypad MP3415 has been tested with all revisions of Bona Light Computers, 

MP3093. 

3.  Numeric Ke

ypad MP3415 may be used with EPS320BT, StimLab™ (Local Touch Screen) and 

StimCor™ products. 

4.  The Keypad may be connected using up to 10 metre extension PS/2 cable, such as is 

contained in MP3403 (StimLab local 10m Cable bundle) and MP3169-10 StimCor Local 10m 
extension cable bundle. 

Note: 

StimLab™ Remote (trunk) Cable, MP3164-17 does not have a provision for connecting 

this Keypad. The StimLab has an on-screen keypad. If you require a hardware keypad at the 
bedside, contact your distributor or Micropace for alternatives. 

 
Cautions 

 

1.  In order to prevent unreliable and unpredictable key operation: 

a)  Use only with Micropace EPS320 software 3.21 or higher. 

b)  Use only in combination with Touch Screen (MP3114) that is enabled in software. 

c)  Do not use concurrently with ke

yboard (either with ‘Y’ splitter or by simultaneous use of 

PS2 and USB ports). 

d)  Do not exceed maximum PS2 cable extension length 10m.  Longer extensions require 

special extender equipment. 

e)  Do not use with a PS2 to USB adaptor, as unexpected actions, including inadvertent 

toggling of pacing, may result. 

Summary of Contents for MP3393

Page 1: ...tructions for Use with the Micropace EPS320 Series of Cardiac Stimulators Micropace Pty Ltd 41 159 Arthur Street Homebush West NSW 2140 Australia www micropace com au Document MP3406 Ver 1 3 April 201...

Page 2: ...e swapped as necessary depending on your needs Note that the three setup functions of naming protocols pace sites and user setups require use of the alpha keys on the Standard Keyboard Compatibility 1...

Page 3: ...he system as normal using the Standard Keyboard Clear any informational messages by pressing ESC until the main stimulator screen is displayed b Press H on the keyboard or the Help menu on the upper r...

Page 4: ...ur Micropace Standard Keyboard and has the following features 1 Dedicated PACE key for easy intuitive usage Press and hold burst pacing is enabled just as in the Standard Keyboard which uses the Space...

Page 5: ...P3393 Page 5 of 9 T_MP3393 Compact Numeric Keypad Leaflet_V1 3 Figure 5 The Keypad Labels Completes numeric entry Increments parameters Exits menus Decrements parameters Numlock toggles numbers vs arr...

Page 6: ...description Re Installation of Standard Keyboard 1 Disconnect the Numeric Keypad PS 2 connector and connect the Standard Keyboard Troubleshooting Symptom Potential Cause Remedy Enter key toggles paci...

Page 7: ...on the product for confirmation FCC CERTIFICATION REQUIRED STATEMENT WARNING This equipment has been certified to comply with the limits for a Class B Computing Device pursuant to Subpart J of Part 15...

Page 8: ...Compact Numeric Keypad MP3393 Page 8 of 9 T_MP3393 Compact Numeric Keypad Leaflet_V1 3...

Page 9: ...pressing ESC _____ Access Help menu and select About the EPS320 _____ Software is verified as 3 21 or above Refer to IFU above for help Install the MP3415 Compact Numeric Keypad _____ Turn off the sy...

Reviews: