5.2
STAINLESS STEEL / TEFLON CONNECTION KIT
The tips and tricks in this section are related to the Fluidic Connect in combination with the
Stainless Steel or Teflon Connection Kit.
I connected all components according to the manual, but the connections still leak. What can be wrong?
1. Make sure the connections are finger tight (see 5.1. Tightening Force), if so DO NOT
tighten the connections any further, it might break the chip.
2. Check if there is no tension on the tubing, causing the tubing to exert a force on the seal.
3. If the tubing exceeds the ferrule edge, this can cause the problem, see figure 5.2.
Fluidic Connect 4515 - User Manual V1.5
A
B
16
F e r u l l e a n d
tube incorrect
assembled
Figure 5.2. Assembly of tubing and ferrule