![Micronics PORTAFLOW 300 Manual Download Page 8](http://html1.mh-extra.com/html/micronics/portaflow-300/portaflow-300_manual_1788003008.webp)
8
PROGRAMING/MAIN MENU
Switch 0n…
Micronics Ltd.
Serial 0000
v
2.00
Press Enter to start
Main Menu
Press SCROLL up or down to move cursor to
required option and press ENTER to select.
Before accessing the flow and data logging
facilities, please ensure that date and time
details are correct (see page 15, Main Menu-
Set-up Portaflow)
Main Menu - Quick Start
Selecting quick start offers the user the easiest
and quickest option to achieve flow
measurement. If the instrument has already
been used, it stores the last application data
entered. This allows the user to read flow on
the same application without spending time
entering new data. Go to ‘Read Flow’ in the
main menu.
If
QUICK START
is selected, proceed with the
following routine. Use the scroll keys to select,
then press ENTER.
QUICK START yy-mm-dd hh:mm:ss
Select the dimension units:
Millimetres
Inches
The instrument now asks for the
Pipe outside
diameter?
After entering the outside diameter
press ENTER.
Pipe wall thickness
now appears on the
display. After entering the pipe wall thickness,
press ENTER.
QUICK START yy-mm-dd hh:mm:ss
Dimension units MILLIMETRES
Pipe outside diameter? 58.0
Pipe wall thickness? 4.0
Pipe lining thickness
now appears on the
display. If the pipe you are measuring has a
lining, enter the
Pipe lining thickness
.
If
nothing is entered the instrument automatically
assumes there is no lining. Press ENTER to
move on. If the application has a lining, enter
the required thickness in the units selected.
Press ENTER to continue.
QUICK START yy-mm-dd hh:mm:ss
Dimension units MILLIMETRES
Pipe outside diameter? 58.0
Pipe wall thickness? 4.0
Pipe lining thickness?
0.0
The instrument now displays
Select pipe wall
material
. By using the scroll keys it is possible
to scroll up or down the options available.
QUICK START yy-mm-dd hh:mm:ss
Select pipe wall material:
Mild Steel
S’ less Steel 316
S’ less Steel 303
Plastic
Cast Iron
Ductile Iron
Copper
Brass
Concrete
Glass
Other (m/s)
Select the required material and press ENTER.
T
he following will only be displayed at this
stage if a lining thickness has been entered.
Use the scroll keys to select the required
material then press ENTER. If
Other
is
selected, enter the propagation rate of the
lining in metres/sec. Contact Micronics if this is
not known.
MAIN MENU yy-mm-dd hh:mm:ss
Quick start
View/Edit Site Data
Select sensor set
Data Logger
Set up RS232
Set up Portaflow
Read flow
QUICK START yy-mm-dd hh:mm:ss
Dimension units
MILLIMETRES
Pipe outside diameter? 58.0