19
DATA LOGGER yy-mm-dd hh:mm:ss
Log name QUICK START
Log data to
MEMORY
Logging interval 5 seconds
START NOW
Start time
97-01-22 00:00:00
Stop time
97-01-25 00:00:00
Memory free 50000
List block names
Next block to view
View log as text
View log as graph
Units l/m
Graph Y axis max. 3450
Clear log
Exit
Log Name
This allows the user to give the data that is
going to be logged, a name. The name will be
displayed at the start of each block of memory
until the instrument has stopped logging.
EDIT LOG NAME yy-mm-dd hh:mm:ss
Use SCROLL to choose, ENTER to select,
•
for SPACE, DELETE to clear, 0 to end
abcdefghijklmnopqrstuvwxyz0123456789
>...........................<
Log Data To
Selecting this option gives the user the choice
of logging to the memory, RS232 or both.
Select the option required by using the scroll
keys and press ENTER (See also
Downloading to Windows pages 12 and 14).
Logging Interval
This option displays a range of times that allow
the user to decide how often the readings need
to be logged. The times range from 5 seconds
to 1 hour. Use the scroll keys to select then
press ENTER.
Start/Stop Now
This starts and stops the logger immediately.
When Start now is displayed press ENTER to
start, the display will change to Stop now.
When Stop now is displayed press ENTER to
stop, the display will change to Start now. This
function defaults the logger to 1 hour of
logging. If a longer period of logging is required
then the
Start/Stop
time
will have to be set
up.
Start/Stop Time
This allows the user to program a time for the
logger to start and stop logging in advance of
going on site. Press ENTER to select and
program as per the instructions for setting time
and date on page 15 – Set up Portaflow.
Note:
Memory free, List block names, Next
block to view, View log as text, View log as
graph, Units, Graph Y axis max, Clear log and
Exit are the same as described on page 10 -
Main Menu - Data Logger
4 - 20mA Output KEY
The 4-20mA can be set to the maximum flow
rate. It is also possible to enter a negative
figure for the minimum output and would
enable a reverse flow to be monitored. The
4mA would then be the maximum reverse flow
(e.g. -100 lpm) and the 20mA would be a
maximum positive flow (e.g. 100 lpm).
mA Out
This displays what the current output is giving
at any particular time.
4 - 20MA yy-mm-dd hh:mm:ss
mA out
0.00
Output
OFF
Units
m/s
Flow at max output
3171
Flow at min output
0.00
Output mA for error
22
Exit
Output
This option allows the user to select between
three different outputs or switching the output
off. The display will read as follows.
Scroll down the options to select required
output, and press ENTER. The display will
then revert back to the
4-20mA
menu and
Flow at max output.
OUTPUT yy-mm-dd hh:mm:ss
OFF
4 - 20mA
0 - 20mA
0 - 16mA
Units
The flow units can be changed at this stage by
selecting them from the keypad. When
selected, scroll down to move onto the next
option.
Flow at Max Output
This sets the output at the top end of the scale
so that the maximum flow gives 20mA (or
16mA).
The instrument automatically defaults to the
maximum flow rate, but by pressing ENTER
the user can scale the output to any level
required. When selected press ENTER to
continue.