AccuPyc II 1340
System Commands
Aug 2013
3-31
Calibration
Press
Alt + 5
to review or edit calibration results; the following prompts display if they were selected
for the calibration. You may edit the information in the prompts if desired
Displays the
Density
or
Volume
(depending on the selection
you made for
Analysis display mode
).
[n]
shown in the prompts to the left represent the cycle num-
ber.
Press
CHOICE
to exclude the displayed density (or volume)
from the calculated average. This will place an asterisk next
to it indicating that it has been excluded. Press
CHOICE
again to remove the asterisk and have it included.
Each time you press
CHOICE
to exclude (or include) the
value, a new deviation is calculated and displayed.
Press
ENTER
to view the values for the next cycle.
Press
SAVE
when you wish to return to the
Reload
prompt.
When you press
SAVE
, all collected data are recalculated
automatically, and all data reduction messages are added
back into the queue.
Displays the insert you selected for the calibration or
None
if an insert was not used. If the selection is incorrect, press
CHOICE
until the correct one is shown, then press
ENTER
.
Choices:
None, 0.1 cm
3
(for 1 cm
3
pycnometer)
None, 3.5 cm
3
, 1.0 cm
3
(for 10 cm
3
pycnometer)
None, 35 cm
3
, 10 cm
3
(for 100 cm
3
pycnometer)
None, 650 cm
3
, 1300 cm
3
(for 2000 cm
3
pycnometer)
Displays the volume of the calibration standard that was
used for the calibration.
Press
ENTER
.
Unit[n]
SN1234
10 cm3
X - X - X
Dn [n] = (density)
Dv [n] = (deviation)
or
Unit[n]
SN1234
10 cm3
X - X - X
V [n] = (volume)
Dv [n] = (deviation
Unit[n]
SN1234 10 cm3
X - X - X
Chamber insert?
(insert)
Unit[n]
SN1234 10 cm3
X - X - X
Volume of Cal Std:
(volume)
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...