Feb 2010
1
Installing the ASAP 2420 1-mmHg Transducer Option
(Units with Serial Number 199 and below)
P/N 242-42805-00
This document contains instructions for installing the 1-mmHg Transducer option only or for
installing both the 1-mmHg transducer and micropore options on an ASAP 2420 analyzer.
Follow the set of instructions that applies to the customer’s system.
Tools Required
The following tools are required for installation of the transducer option:
•
1/2-in. wrench
•
9/16-in. wrench
•
3/4-in. wrench
•
Phillips screwdriver
•
Small flat-blade screwdriver
Installing the 1-mmHg Transducer Option Only
This set of instructions describes how to install the 1-mmHg Transducer option on an ASAP
2420 system equipped with the Micropore option.
1. Backfill the manifold to 760 mmHg.
2. Close the ASAP 2420 software.
3. Turn off power to the vacuum pumps and to the analyzer.
4. Remove the four retaining screws from the top panel of the ASAP 2420.