
13
STANDARD CLEANING OF BURETTE:
Whenever cleaning is required, run the burette under distilled water.
a.
c.
e.
b.
d.
f.
Fill the barrel completely with distilled water and press the fast dispense
button to dispense the water completely.
Place a suitable receiving vessel below the opening of the discharge
tube and rinse the instrument several times by completely filling and
emptying it.
Set the valve to recirculation and rinse the instrument several times by
completely filling and emptying it.
This process can optionally be repeated with a suitable cleaning agent.
Unscrew the chuck nut by rotating it in anticlockwise direction and pull
out the delivery pipe. Clean the delivery pipe with deionized water and
assemble it again.
Next, rinse the burette again with deionized water.
INTENSIVE CLEANING OF BURETTE:
The dispensing cylinder, valve, telescoping filling tube and titrating tube
are filled with reagent. For this reason, always perform a standard
cleaning before dismantling the instrument. In order to avoid confusion
about the components, do not dismantle more than one instrument at a
time. A calibration and any necessary adjustment, must be carried out
after dismantling or replacement of a piston/cylinder assembly.
Note: Use rubbing alcohol on a cloth or cotton to clean the external
body.
Proper storage of digital burette is as important as cleaning and
calibrating the instrument. The digital burette and its all accessories
must be stored in a clean, cool and dry place. The storage temperature
should be ranging from –20 °C to +50 °C (from –4 °F to 122 °F) is advised
with relative humidity between 5% and 95%. Another point to be
remembered is that the instrument should be stored in an upright
position as recommended by the manufacturer.