E-burette is commonly used in the titration process to determine the
concentration of an unknown solution. Typically, titrant (known solution) is
added from a burette to a known volume of analyte (unknown solution)
until the reaction is complete. An indicator is used to find the endpoint of
the titration.
10
Click the purging mode from the home screen on the control panel.
The screen on the control panel will warn the user to turn the knob
to recirculation mode.
Turn the knob to recirculation mode and click the tick mark on the
control panel screen when ready.
The digital burette will purge automatically. If the instrument is still
not properly purged, repeat the purging process again till no large
air bubble is visible below the piston.
BEFORE STARTING USING THE DIGITAL BURETTE,
IT IS RECOMMENDED TO ENSURE BUBBLE FREE DISPENSING:
Remove the cap from the discharge tube.
Click on the titration mode on the home screen. The screen on the
control panel will warn the user to turn the knob to titration mode.
Turn the knob clockwise to set at titration mode and click the tick
mark on the screen of the control panel when ready.
ONCE THE E-BURETTE IS PROPERLY SETUP,
FOLLOW THE STEPS BELOW TO PERFORM THE TITRATION PROCESS.