Troubleshooting

16.

Guidelines for Return

Follow these guidelines, before sending the instrument for repair:

Please clean the instrument carefully. 

On no account should the sent instrument contain any reagent.

The sent instrument can only be checked and repaired if it has 

been carefully cleaned and decontaminated, beforehand.

Forcefully opening the instrument invalidates any warranty claim.

While packing the instrument for return, please enclose a detailed 

description of the trouble that has occurred and the reagents that 

were used with the instrument.

17.

28

27

Problem

Possible cause

Solution

Piston difficult to move

Formation of crystals or 

deposition of dirt

Stop dispensing immediately.

Loosen piston in circular motion, 

but do not disassemble. 

Follow the cleaning instructions. 

(Refer to section13)

Air bubbles appear 

in the instrument

Reagent with high vapour 

pressure has been drawn 

in too quickly.

The instrument has not 

been primed.

Filling tube is loose or damaged.

Liquid reservoir is empty.

The filling action is too fast.

The piston is leaking.

The discharge valve is leaking

Draw the reagent slowly.

Prime   the instrument 

(Refer to section 9)

Push the filling tube on firmly.

If necessary, cut off approx. 

1 cm of the tube at the upper 

end and then re-connect it 

or replace it.

Refill reservoir and 

prime the unit.

Fill and dispense slowly.

Clean the piston. Refer to 

section 13. If problem persists, 

replace it.

Clean by flushing thoroughly

with distilled water. 

(Refer to section 13)

Dispensing not possible

Dispensing nozzle is blocked.

Discharge valve is stuck.

Disassemble the dispensing 

nozzle and flush thoroughly 

with distilled water.

Clean the unit by immersing 

valve assembly in distilled water. 

(Refer to section 13)

Wrong dispensing volume

The instrument is not calibrated. Follow user calibration steps.

(Refer to section 12)

Barrel does not fill with liquid

Inlet tube is not fitted firmly.

Connect inlet tube correctly. 

(Refer to section 8)

Filling not possible

Adjust volume to 

minimum setting.

Set to the required volume. 

(Refer to section 10)

Summary of Contents for BEATUS

Page 1: ... Kursi Road Lucknow 226026 UP India Call at 91 8081461996 91 9918625629 For queries info microlit com For feedback customercare microlit com www microlit com OPERATION MANUAL Bottle Top Dispenser with Recirculation Valve and Springless Valve Technology ...

Page 2: ...ecommended Reagents First Steps Setting Up Priming Volume Setting and Dispensing Error Limits User Calibration Procedure Maintenance Cleaning Disassembling and Assembling for Cleaning and Servicing Autoclaving Troubleshooting Guidelines for Return 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 1 1 2 3 3 4 11 13 15 16 18 19 20 21 24 27 28 ...

Page 3: ...her person Only dispense into appropriate vessels and avoid splashes Do not press the piston when the discharge tube closure is attached Do not remove the discharge tube while the dispensing cylinder is being filled Reagents can accumulate in the cap of the discharge tube Thus clean it regularly Do not carry the mounted instrument by the cylinder sleeve or the valve block Breakage or loosening of t...

Page 4: ...plosive liquids e g Carbon Disulfide Suspensions e g of charcoal as the solid particles may clog or damage the instrument Liquids that attack PP cap The permissible concentration of dissolved Sodium Azide is not more than 0 1 4 Storage Conditions Store the instrument and accessories in a clean cool and dry place Storage temperature may vary from 20 C to 50 C from 4 F to 122 F 5 Operating limitation...

Page 5: ... B 2 A A B 2 A B 2 A B 2 B 2 A C 2 B 2 A A A B 2 B 2 B 2 B 2 A A A A List of Reagents Dimethylacetamide Covi Ox T70 Mixed Tocopherol B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 1 2 B 1 2 C 1 2 A A A A A A A A A A A A A A A A A A Trichloromethane ...

Page 6: ... B 2 B 2 B 2 B 1 C 3 C 2 C 2 A A A A A A A A A A A A A A A A A A A A A A A A A A List of Reagents Oleum Fuming Sulfuric acid Potassium dihydrogen phosphate B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 Picric acid C 2 A A A A A A A A A A A A A A A A A A A A A A A A A A ...

Page 7: ...le Cap Volume Adjustment Knob Middle Body Barrel Adapter Bottle Adapter Telescopic Tube List of Reagents P B 2 Triethylamine B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 B 2 Sulfuric acid 10 Sulfuric acid 10 75 Sulfuric acid Cold conc Sulfuric acid Hot conc A B 1 A B 2 A A A A A A A A A A A A A A A A A A A A A ...

Page 8: ...kage Verify that the package includes 1 Operation Manual 7 1 2 3 1 Bottle Top Dispenser Fig 7 1 1 Telescopic Tube Fig 7 2 1 Calibration Tool Fig 7 3 Fig 7 1 4 5 6 5 Bottle Adapters 28 mm 32 mm 38 mm 40 mm 45 mm Fig 7 4 1 Calibration Certificate Fig 7 5 Fig 7 4 Fig 7 5 Fig 7 2 Fig 7 3 ...

Page 9: ...press the piston down when the cap is on Avoid splashing the reagent The reagent can drip out from the discharge tube and cap Fig 8 6 Fig 8 1 Setting Up Wear protective clothing protective gear for the eyes and hands Follow all safety instructions and observe limitations of use and operating constraints Refer to sections 2 3 4 8 1 2 3 Fig 8 2 Adjust the length of the telescopic tube to fit your par...

Page 10: ...insed carefully and discard the first few samples dispensed Avoid splashing For priming gently pull the piston up approximately by 30mm and then down rapidly till the lower end Fig 9 3 Repeat this procedure five times Fig 10 1 Fig 10 2 Volume Setting and Dispensing Volume setting Volume Adjustment Knob is simple and easy to operate There are two positions of the knob as shown To set the volume follo...

Page 11: ...e well within the limits of DIN EN ISO 8655 5 11 Remove cap from the discharge tube Fig 10 3 Before using the instrument turn the valve to Dispensing Fig 10 4 Hold the discharge tube orifice against the inner wall of the appropriate receiving vessel Fig 10 5 Fig 10 6 Fig 10 7 Wipe off the discharge tube against the inner wall of the receiving vessel 4 5 6 Gently lift the piston until the upper end ...

Page 12: ...toclaving Prior to changing the valve Regularly when using liquids which form deposits e g crystallizing liquids Regularly when liquids accumulate in the cap All maintenance cleaning procedures should be carried out while wearing suitable eye protection and protective clothing When in doubt consult your safety officer Make sure that the instrument is completely empty Place the instrument into an em...

Page 13: ...nut with the help of calibration tool to disassemble the piston and shaft from the main housing Fig 14 2 Clean the cylinder with a bottle brush If necessary carefully remove deposits at the edge of the glass cylinder Fig 14 5 Insert the piston completely into the cylinder and then reassemble the instrument using the calibration tool by screwing back the piston Fig 14 6 Snap the cap back to complet...

Page 14: ...emble the Delivery Pipe Fig 14 9 Fig 14 11 Fig 14 8 Fig 14 10 Fig 15 2 Fig 15 3 Straighten the delivery pipe Fig 14 8 Pull the cap outwards to expose the calibration nut Fig 15 1 Unscrew the calibration nut with the help of calibration tool to disassemble the piston and shaft from the main housing Fig 15 2 After unscrewing pull out the shaft Fig 15 3 Unscrew the chuck nut and pull out the delivery...

Page 15: ...p the cap back to complete the assembly Fig 15 7 Cooling After autoclaving cool the dispenser at room temperature for at least 2 hours before use Re assembling after autoclaving 4 5 1 Fig 15 4 Fig 15 5 This is the piston shaft sub assembly Fig 15 4 Autoclave the two sub assemblies at 121 C and 15 psi pressure for 10 15 minutes Fig 15 5 6 The volume adjustment knob should always be kept in the unlo...

Page 16: ...ent has not been primed Filling tube is loose or damaged Liquid reservoir is empty The filling action is too fast The piston is leaking The discharge valve is leaking Draw the reagent slowly Prime the instrument Refer to section 9 Push the filling tube on firmly If necessary cut off approx 1 cm of the tube at the upper end and then re connect it or replace it Refill reservoir and prime the unit Fill a...

Reviews: