Dispensing
Wear protective clothing, protective gear for the eyes and hands.
Liquid may accumulate in the cap. To avoid splashes, dispense slowly.
Follow all safety instructions and observe limitations of use and
operating constraints.
1.
2.
3.
Error Limits
The error limits (Accuracy and Coefficient of Variation) mentioned below are
in accordance with the nominal capacity (or maximum volume) indicated on
the instrument. These are obtained by using the instrument with distilled water
at equilibrium, ambient temperature of 20 °C or 68°F, while operating the device
smoothly and steadily.
The error limits are well within the limits of DIN EN ISO 8655-5.
11.
Remove cap from the discharge tube.
(Fig. 10.3)
Before using the instrument, turn the valve
to ‘Dispensing’.
(Fig. 10.4)
Hold the discharge tube orifice against the
inner wall of the appropriate receiving vessel.
(Fig. 10.5)
Fig.
10.6
Fig.
10.7
Wipe off the discharge tube against the inner wall of the receiving vessel.
4.
5.
6.
Gently lift the piston until the upper end. Press the
piston downwards slowly and steadily with minimal
force until the lower end.
(Fig. 10.6)
Reattach the cap to the discharge tube.
(Fig. 10.7)
Volume
Range
0.25-2.5 ml
0.05 ml
0.6
0.2
0.2
0.2
0.2
0.2
0.2
0.005
0.010
0.020
0.060
0.120
0.200
0.015
0.030
0.060
0.180
0.360
0.600
0.6
0.6
0.6
0.6
0.6
0.1 ml
0.2 ml
0.5 ml
1.0 ml
2.0 ml
0.5-5 ml
1-10 ml
2.5-30 ml
5-60 ml
10-100 ml
Increment
Specifications ISO 8655
Accuracy
±% ±ml
CV
±% ±ml