Fig.
15.6
26
25
Dispenser is now ready for use. Recalibration is required after autoclaving.
2.
Fig.
15.7
Insert the piston completely into the cylinder and
then reassemble the instrument.
Use the calibration tool to screwing the piston.
(Fig. 15.6)
Snap the cap back to complete the assembly.
(Fig. 15.7)
Cooling
After autoclaving, cool the dispenser at room temperature for
at least 2 hours before use.
Re-assembling after autoclaving
4.
5.
1.
Fig.
15.4
Fig.
15.5
This is the piston-shaft sub-assembly.
(Fig. 15.4)
Autoclave the two sub-assemblies at 121°C and
15 psi pressure for 10-15 minutes.
(Fig. 15.5)
6. The volume adjustment knob should always be kept in the 'unlocked' position while
autoclaving (Refer fig 10.2).