Microlife FR1DQ1-B User Manual Download Page 12

 

  If no message showed on App to confirm which patient log the data should be assigned to, please record the temperature data 

showed on FR1DQ1-B into the App manually or takes a new measurement and start Transfer Data step from 1 to 6 again. 

Bluetooth® Icon Indicator 

 

The Bluetooth® icon indicator on your device, located in the middle right portion of the display, is designed to provide 

information about the connection between your device and smartphone.   

 

The Bluetooth® function is enabled: The Bluetooth® icon indicator

14

 

appears steady on the display. 

 

 

Connecting with a device or uploading the data: The Bluetooth® icon indicator

14

 flashes. 

 

 

The Bluetooth® function is automatically switched off during measurement; after measurement the device will switch 

it on and upload the data to the «Microlife Connected » App.

   

NOTES 

 

The Bluetooth® of this thermometer is a  medical device data system  (MDDS) as its only function is for records and no 

additional functions. It does not modify the data or modify the display of the data, and it does not by itself control the 

functions or parameters of any other medical device. It  is not active when the thermometer is recording data or during 

measurement. The thermometer will not sound any alarm with or without Bluetooth®. The Bluetooth® is used only to 

transfer data from point A to point B. The App on your smart devices cannot be used to start or stop the thermometer, nor 

update the firmware of thermometer via Bluetooth®.   

Wireless communication interference 

 

This product operates in an unlicensed ISM band at 2.4 GHz. In the event this product is used near other wireless devices 

such as microwave and wireless LAN, which operate on the same frequency band as this product, there is a possibility that 

interference may occur. If interference occurs, stop the operation of the other devices or relocate this product away from 

other wireless devices before attempting to use it. 

 

11. Error Messages 

 

Measured temperature too high

18 : 

Displays «

ErrH

» when measured temperature is higher than 43°C/109.4°F in body mode or 

Displays «

H

» when measured temperature is higher than 99.9°C /211.9°F in object mode. 

 

Measured temperature too low

19 : 

Displays «

ErrL

» when measured temperature is lower than 34.0 °C / 93.2 °F in body mode 

or Displays «

L

» when measured temperature is lower than 0.1 °C / 32.2 °F in object 

mode

 

Ambient temperature too high

20 : 

Displays «

AH

» when ambient temperature is higher than 40.0 °C / 104.0°F

 

in body model 

and object model.   

 

Ambient temperature too low

21 : 

Displays «

AL

» when ambient temperature is lower than 15.0 °C / 59.0 °F in body mode and 

5.0 °C / 41.0 °F in object mode.   

 

Error function display

22 : 

«

Err

»: The system has a malfunction. 

«

Errd

»: Device is over 5cm from forehead /object. 

«

Er2

»: Keep the measuring distance of 1-5 cm. 

Do not touch the bottom side (sensing area) of the measuring sensor. 

 

Blank display

23 : 

Check if the batteries have been inserted correctly. Also check polarity (<+> and <->) of the 

batteries. 

Summary of Contents for FR1DQ1-B

Page 1: ...FR1DQ1 B Infrared Forehead Thermometer Distributed by Microlife USA Inc 1617 Gulf to Bay Blvd Floor 2 Clearwater FL 33755 UNITED STATES Phone 727 442 5353 www microlifeusa com Made in China...

Page 2: ...8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27...

Page 3: ...temperature too high 21 Ambient temperature too low 22 Error function display 23 Blank display 24 Flat battery 25 Date Time 26 Beeper function setting 27 Protective cap Read the instructions carefully...

Page 4: ...Connected Health APP The memory data can be transferred to the smart phone via Bluetooth Table of Contents 1 The Advantages of this Thermometer Measures in a matter of seconds Measurement with distanc...

Page 5: ...urface temperature in a baby s bottle Surface temperature of a baby s bath Accurate and reliable The unique probe assembly construction incorporates an advanced infrared sensor ensuring that each meas...

Page 6: ...re measurement the thermometer and the person whose temperature is to be measured should be in the room temperature for at least 30 minutes If the measurement result is not consistent with the patient...

Page 7: ...ice detect the distance is appropriate within 5cm the screen shows good 10 with 1 short beep then press START IO button 4 to start measuring For 1s mode when the device detects the distance is appropr...

Page 8: ...y mode directly If there is any item of time setting have not been completed they will be show with no information Setting the beeper 26 At any status press MODE 6 button for 3 seconds will enter beep...

Page 9: ...ify the completion of measurement 7 Read the recorded temperature from the LCD display 8 For the next measurement remove the thermometer from the forehead and wait until the C F icon is flashing Follo...

Page 10: ...s of age with a compromised immune system and for whom the presence or absence of fever is critical 2 When the user is learning how to use the thermometer for the first time until he she has familiari...

Page 11: ...enter My profile 3 Edit profile information then select Save to complete Profile setup Transfer Data 1 Turn on Bluetooth of Smartphone 2 Open the app on the Smartphone and select the Temperature Overv...

Page 12: ...ot be used to start or stop the thermometer nor update the firmware of thermometer via Bluetooth Wireless communication interference This product operates in an unlicensed ISM band at 2 4 GHz In the e...

Page 13: ...devices must be disposed of in accordance with the locally applicable regulations not with domestic waste 14 Guarantee This device is covered by 5 years guarantee from the date of purchase The guarant...

Page 14: ...5 40 0 C 41 104 0 F 15 95 relative maximum humidity Storing conditions 25 55 C 13 131 F 15 95 relative maximum Automatic Switch off Approx 30 sec when power on without measurement after measurement Ba...

Page 15: ...ation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interf...

Page 16: ...ronment guidance RFemissions CISPR 11 Group 1 The model FR1DQ1 B uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in n...

Page 17: ...l or hospital environment Surge IEC 61000 4 5 1 kV line s and neutral N A Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltag...

Page 18: ...he recommended separation distance calculated from the equation applicable to the frequency of the transmitter Conducted RF IEC 61000 4 6 3 Vrms 150 kHz to 80 MHz N A Recommended separation distance R...

Page 19: ...TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured...

Page 20: ...t power of transmitter Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 23 0 1 0 38 0 73 1 1 2 2 3 10 3 8 7 3 100 12 23 Fo...

Reviews: