Before and after each treatment

Disassemble the nebuliser 

5

 by turning the top counterclockwise 

and remove the medicine conduction cone. Wash the components 
of the disassembled nebuliser, the mouthpiece 

7

 and the nose 

piece 

AL

 by using tap water; dip in boiling water for 5 minutes. 

Reassemble the nebuliser components and connect it to the air 
tube connector, switch the device on and let it work for 10-15 
minutes. 

Wash masks and air tube with warm water.

Only use cold disinfecting liquids following the manufac-
turer’s instructions.

Do not boil nor autoclave the air tube and masks.

5. Maintenance, Care, and Service

Order all spare parts from your dealer or pharmacist, or contact 
Microlife-Service (see foreword).

Replacement of the nebuliser

Replace the nebuliser 

5

 after a long period of inactivity, in cases 

where it shows deformities, breakage, or when the vaporiser head 

5

-a is obstructed by dry medicine, dust, etc. We recommend to 

replace the nebuliser after a period between 6 months and 1 year 
depending on the usage.

Only use original nebulisers!

Replacement of the air filter

In normal conditions of use, the air filter 

AM

 must be replaced 

approximately after 200 working hours or after each year. We 
recommend to periodically check the air filter (10 - 12 treatments) 
and if the filter shows a grey or brown colour or is wet, replace it. 
Extract the filter and replace it with a new one. 

Do not try to clean the filter for reusing it.

The air filter shall not be serviced or maintained while in use 
with a patient.

Only use original filters! Do not use the device without filter!

6. Malfunctions and Actions to take

The device cannot be switched on

Ensure the power lead 

2

 is correctly plugged into the socket.

Ensure the ON/OFF switch 

3

 is in the position «

I

».

Make sure that the device has been operating within operating 
limits indicated in this manual (30 min On / 30 min Off).

The nebuliser functions poorly or not at all

Ensure the air tube 

6

 is correctly connected at both ends.

Ensure the air tube is not squashed, bent, dirty or blocked. If 
necessary, replace with a new one.

Ensure the nebuliser 

5

 is fully assembled and the vaporiser 

head 

5

-a is placed correctly and not obstructed.

Ensure the required medication has been added.

7. Guarantee

This device is covered by a 

5 year guarantee

 from the date of 

purchase. During this guarantee period, at our discretion, Microlife 
will repair or replace the defective product free of charge.
Opening or altering the device invalidates the guarantee.
The following items are excluded from the guarantee:

Transport costs and risks of transport.

Damage caused by incorrect application or non-compliance 
with the instructions for use.

Damage caused by leaking batteries.

Damage caused by accident or misuse.

Packaging/storage material and instructions for use.

Regular checks and maintenance (calibration).

Accessories and wearing parts: Nebuliser, masks, mouthpiece, 
nose piece, tube, filters, nasal washer (optional).

Should guarantee service be required, please contact the dealer 
from where the product was purchased, or your local Microlife 
service. You may contact your local Microlife service through our 
website: 
www.microlife.com/support
Compensation is limited to the value of the product. The guarantee 
will be granted if the complete product is returned with the original 
invoice. Repair or replacement within guarantee does not prolong 
or renew the guarantee period. The legal claims and rights of 
consumers are not limited by this guarantee. 

Summary of Contents for BR-CN188

Page 1: ...angZai Road Xinqiao Street Bao An District 518125 Shenzhen City Guangdong Province PRC Phone 86 755 27464589 www bi rich com SUNGO Cert GmbH Lindenstra e 48 52 40233 D sseldorf Germany Phone 49 211 97...

Page 2: ...o Meno z kazn ka Ime in priimek kupca Ime i prezime kupca V s rl neve Ime i prezime kupca Serial Number Num r de serie Numer seryjny V robn slo V robn slo Serijska tevika Serijski broj Sorozatsz m Ser...

Page 3: ...rolife AG Table of Contents 1 Explanation of Symbols 2 Important Safety Instructions 3 Preparation and Usage of this Device 4 Cleaning and Disinfecting Cleaning and disinfecting of the accessories Bef...

Page 4: ...cerning the electrical devices and in particular Never touch the device with wet or moist hands Place the device on a stable and horizontal surface during its operation Do not pull the power cord or t...

Page 5: ...described in the section Cleaning and Disinfecting 1 Assemble the nebuliser kit AK Ensure that all parts are complete 2 Fill the nebuliser with the inhalation solution as per your doctor s instruction...

Page 6: ...switched on Ensure the power lead 2 is correctly plugged into the socket Ensure the ON OFF switch 3 is in the position I Make sure that the device has been operating within operating limits indicated...

Page 7: ...ate 0 13 ml min Percentage of fill volume emitted per min 3 5 Residual Volume 0 5 ml Particle size MMAD 2 08 m GSD geometric stan dard deviation 1 87 m RF respirable fraction 0 5 5 m 95 8 Large partic...

Page 8: ...6...

Page 9: ...i 2 Wa ne wskaz wki dotycz ce bezpiecze stwa 3 Przygotowanie urz dzenia do pracy oraz jego u ywanie 4 Czyszczenie i dezynfekcja Czyszczenie i dezynfekcja akcesori w Przed i po ka dym zabiegu 5 Przecho...

Page 10: ...przestrzeganie wskaz wek dotycz cych przechowywania i u ytkowania zamieszczonych w cz ci Specyfikacja techniczna Chro urz dzenie przed wod i wilgoci ekstremalnymi temperaturami wstrz sami i upadkiem...

Page 11: ...jenta Urz dzenie natomiast jest przeznaczone do u ytku przez wielu pacjent w Nie nale y przechyla inhalatora pod k tem przekraczaj cym 60 Nie u ywaj urz dzenia w pobli u wyst powania silnego pola elek...

Page 12: ...as nie by u ywany w przypadku kiedy jest uszkodzony zdeformowany p kni ty lub 5 a jest zaschni ty przez resztki leku kurz Zale camy wymieni nebulizator po 6 miesi cach lub po 1 roku w zale no ci od cz...

Page 13: ...nym Nebulizator ustnik i maski s cz ciami typu BF i Dane techniczne mog ulec zmianie bez uprzedniego powiadomienia Prosimy o zg aszanie wszelkich powa nych incydent w kt re mia y miejsce w zwi zku z u...

Page 14: ...12 dzia aniem niepo danym do w a ciwych w adz lokalnych produ centa lub do autoryzowanego przedstawiciela w Europie EC REP...

Reviews: