BP B3 BT

EN

Permanently high blood pressure values can damage your 

health and must be treated by your doctor!

 

Always discuss your values with your doctor and tell him/her if 

you have noticed anything unusual or feel unsure. 

Never rely 

on single blood pressure readings. 

Under no circumstances should you alter the dosages of 

drugs or initiate a treatment without consulting your doctor.

Deviations

 between measurements taken by your doctor or in 

the pharmacy and those taken at home are quite normal, as 

these situations are completely different.

The pulse display is not suitable for checking the 

frequency of heart pacemakers!

If you are 

pregnant

, you should monitor your blood pressure 

regularly as it can change drastically during this time.

Ensure that children do not use this device unsupervised; 

some parts are small enough to be swallowed. Be aware of 

the risk of strangulation in case this device is supplied with 

cables or tubes.

Device care

Clean the device only with a soft, dry cloth.

Cleaning the cuff

The cuff delivered with this device is washable.

1. Remove the cuff connector 

AT

 from the cuff tube 

AK

 and care-

fully pull the bladder through the opening at the edge of the cuff 

cover.

2. Hand wash the cuff cover in soapsuds: not hotter than 30 °C.

3. Completely dry the cuff cover by linen drying.

4. Loop the cuff tube back through its opening and carefully place 

the bladder flat in the cuff cover.

5. Reattach the cuff connector on the cuff tube.

The bladder must lay straight in the cuff cover, not folded.

Do not use fabric softener.

WARNING: 

Do not wash the cuff in a washing machine or 

dishwasher!

WARNING: 

Do not dry the cuff cover in a tumble dryer!

WARNING: 

Under no circumstances may you wash the 

inner bladder!

Accuracy test

We recommend this device is tested for accuracy every 2 years or 

after mechanical impact (e.g. being dropped). Please contact your 

local Microlife-Service to arrange the test (see foreword).

Disposal

Batteries and electronic devices must be disposed of in 

accordance with the locally applicable regulations, not with 

domestic waste. 

10. Guarantee

This device is covered by a 

5 year guarantee

 from the date of 

purchase. The guarantee is valid only on presentation of the guar-

antee card completed by the dealer (see back) confirming date of 

purchase or the receipt.

Batteries and parts that become worn with use are not included.

Opening or altering the device invalidates the guarantee.

The guarantee does not cover damage caused by improper 

handling, discharged batteries, accidents or non-compliance 

with the operating instructions.

The cuff has a functional guarantee (bladder tightness) for 2 years.

Please contact your local Microlife-Service (see foreword).

11. Technical Specifications

Operating conditions:

10 - 40 °C / 50 - 104 °F

15 - 95 % relative maximum humidity

Storage conditions:

-20 - +55 °C / -4 - +131 °F

15 - 95 % relative maximum humidity

Weight:

402 g (including batteries)

Dimensions:

138 x 94.5 x 62.5 mm

Measuring procedure:

oscillometric, corresponding to Korotkoff 

method: Phase I systolic, Phase V diastolic

Measurement range:

20 - 280 mmHg – blood pressure

40 - 200 beats per minute – pulse

Cuff pressure 

display range:

0 - 299 mmHg

Resolution:

1 mmHg

Static accuracy:

pressure within ± 3 mmHg

Pulse accuracy:

± 5 % of the readout value

Voltage source:

4 x 1.5 V alkaline batteries; size AA

Mains adapter DC 6V, 600 mA (optional)

Battery lifetime:

approx. 920 measurements 

(using new batteries)

IP Class:

IP20

Reference to 

standards:

EN 1060-1 /-3 /-4; IEC 60601-1;

IEC 60601-1-2 (EMC); IEC 60601-1-11

Summary of Contents for BPB3 BT

Page 1: ...yhnite sa tesnému alebo priliehavému odevu na hornej časti ramena AIzogibajte se tesnim oblačilom Izbegavajte usku odeću na nadlaktici Kerülje a vastag vagy szoros ruhákat a felkaron Izbjegavajte usku odjeću na nadlaktici Place the artery mark on the cuff over your artery Umieść mankiecie na tętnicy Поместите манжету так чтобы значок артерии на манжете находился над артерией руки в сгибе локтя Пос...

Page 2: ...n Vyhněte se jídlu koupání kouření nebo kofeinu přibližně 30 minut Vyhnite sa jedlu kúpaniu fajčeniu alebo kofeínu približne 30 minút Izogibajte se hrani kopanju in kajenju približno 30 minut Izbegavajte hranu kupanje pušenje i kofein približno 30 minuta Kerülje az evést fürdést dohányzást vagy koffeinfogyasztást a megelőző kb 30 percben Izbjegavajte jesti kupati se pušiti i kofein oko 30 min Avoi...

Page 3: ...nfor mation on our products Stay healthy Microlife AG This device uses the same measuring technology as the award winning BP 3BTO A model tested according to the British and Irish Hypertension Society BIHS protocol Table of Contents 1 Using the Device for the First Time Inserting the batteries Setting the date and time Selecting the correct cuff Selecting the user Selecting standard or MAM mode 2 ...

Page 4: ...t the minutes and pressed the MAM button the date and time are set and the time is displayed 5 If you want to change the date and time press and hold the MAM button for approx 3 seconds until the year number starts to flash Now you can enter the new values as described above Selecting the correct cuff Microlife offers different cuff sizes Select the cuff size to match the circumference of your upp...

Page 5: ...The cuff will now pump up automatically Relax do not move and do not tense your arm muscles until the measurement result is displayed Breathe normally and do not talk 4 The cuff fit check AR on the display indicates that the cuff is perfectly placed If the icon BT A appears the cuff is fitted suboptimally but it is still ok to measure 5 When the correct pressure is reached the pumping stops and th...

Page 6: ...elect either user 1 or 2 by pressing the user button 5 Viewing the average of the last 28 days Press the M button 3 briefly when the device is switched off The display first shows M BP and 28A which stands for the average of measurement values of the last 28 days Blood pressure readings with suboptimal cuff fit BT A are not considered in the average value Viewing the clinical blood pressure averag...

Page 7: ...alues although date and time must be reset the year number therefore flashes automatically after the batteries are replaced Which batteries and which procedure Use 4 new long life 1 5 V size AA alkaline batteries Do not use batteries beyond their date of expiry Remove batteries if the device is not going to be used for a prolonged period Using rechargeable batteries You can also operate this devic...

Page 8: ...e See also chapter Appearance of the Irregular Heartbeat IHB Symbol of this user manual The measurement results given by this device is not a diag nosis It is not replacing the need for the consultation of a physi cian especially if not matching the patient s symptoms Do not rely on the measurement result only always consider other potentially occurring symptoms and the patient s feedback Calling ...

Page 9: ...over in a tumble dryer WARNING Under no circumstances may you wash the inner bladder Accuracy test We recommend this device is tested for accuracy every 2 years or after mechanical impact e g being dropped Please contact your local Microlife Service to arrange the test see foreword Disposal Batteries and electronic devices must be disposed of in accordance with the locally applicable regulations n...

Page 10: ...s reserved The Bluetooth word mark and logos are registered trademarks owned by Bluetooth SIG Inc and any use of such marks by Micro life Corp is under license Other trademarks and trade names are those of their respective owners Expected service life Device 5 years or 10000 measurements Accessories 2 years ...

Reviews: