BP B3 BT

EN

The bottom, right hand section of the display shows a 1, 2 or 3 to 

indicate which of the 3 measurements is currently being taken.

There is a break of 15 seconds between the measurements. 

A count down indicates the remaining time.

The individual results are not displayed. Your blood pressure will 

only be displayed after all 3 measurements are taken.

Do not remove the cuff between measurements.

If one of the individual measurements was questionable, a fourth 

one is automatically taken.

2. Checklist for Taking a Reliable Measurement

Avoid activity, eating or smoking immediately before the 

measurement.

Sit down on a back-supported chair and relax for 5 minutes. 

Keep the feet flat on the floor and do not cross your legs.

Always measure on the same arm

 (normally left). It is recom-

mended that doctors perform double arm measurements on a 

patients first visit in order to determine which arm to measure in 

the future. The arm with the higher blood pressure should be 

measured.

Remove close-fitting garments from the upper arm. To avoid 

constriction, shirt sleeves should not be rolled up - they do not 

interfere with the cuff if they are laid flat.

Always ensure that the correct cuff size is used (marking on the 

cuff).

Fit the cuff closely, but not too tight.

Make sure that the cuff is positioned 1-2 cm above the elbow.

The 

artery mark

 on the cuff (ca. 3 cm long bar) must lie over 

the artery which runs down the inner side of the arm.

Support your arm so it is relaxed.

Ensure that the cuff is at the same height as your heart.

3. Taking a Blood Pressure Measurement

1. Select standard (single measurement) or MAM mode (automatic 

triple measurement): see details in chapter «1.».

2. Press the ON/OFF button 

1

 to start the measurement.

3. The cuff will now pump up automatically. Relax, do not move 

and do not tense your arm muscles until the measurement 

result is displayed. Breathe normally and do not talk.

4. The cuff fit check 

AR

 on the display indicates that the cuff is 

perfectly placed. If the icon 

BT

-A appears, the cuff is fitted 

suboptimally, but it is still ok to measure.

5. When the correct pressure is reached, the pumping stops and the 

pressure falls gradually. If the required pressure was not reached, 

the device will automatically pump some more air into the cuff.

6. During the measurement, the pulse indicator 

BM

 flashes in the 

display.

7. The result, comprising the systolic 

AM

 and the diastolic 

AN

 blood 

pressure and the pulse rate 

AO

 is displayed. Note also the 

explanations on further display symbols in this booklet.

8. When the device has finished measuring, remove the cuff.

9. Switch off the device. (The monitor does switch off automatically 

after approx. 1 min.).

You can stop the measurement at any time by pressing the 

ON/OFF button (e.g. if you feel uneasy or an unpleasant 

pressure sensation). 

This monitor is specially tested for use in pregnancy and 

pre-eclampsia. When you detect unusual high readings in 

pregnancy, you should measure after a short while again 

(eg. 1 hour). If the reading is still too high, consult your 

doctor or gynecologist.

Manual inflation

In case of high systolic blood pressure (e.g. above 135 mmHg)

it can be an advantage to set the pressure individually. Press the ON/

OFF button after the monitor has been pumped up to a level of 

approx. 30 mmHg (shown on the display). Keep the button pressed 

until the pressure is about 40 mmHg above the expected systolic 

value – then release the button. 

How not to store a reading

As soon as the reading is displayed press and hold the ON/OFF 

button 

1

 until «

M

» 

BP

 is flashing. Confirm to delete the reading by 

pressing the MAM button 

4

.

«

CL

» is displayed when the reading is deleted from the 

memory successfully.

Summary of Contents for BPB3 BT

Page 1: ...yhnite sa tesnému alebo priliehavému odevu na hornej časti ramena AIzogibajte se tesnim oblačilom Izbegavajte usku odeću na nadlaktici Kerülje a vastag vagy szoros ruhákat a felkaron Izbjegavajte usku odjeću na nadlaktici Place the artery mark on the cuff over your artery Umieść mankiecie na tętnicy Поместите манжету так чтобы значок артерии на манжете находился над артерией руки в сгибе локтя Пос...

Page 2: ...n Vyhněte se jídlu koupání kouření nebo kofeinu přibližně 30 minut Vyhnite sa jedlu kúpaniu fajčeniu alebo kofeínu približne 30 minút Izogibajte se hrani kopanju in kajenju približno 30 minut Izbegavajte hranu kupanje pušenje i kofein približno 30 minuta Kerülje az evést fürdést dohányzást vagy koffeinfogyasztást a megelőző kb 30 percben Izbjegavajte jesti kupati se pušiti i kofein oko 30 min Avoi...

Page 3: ...nfor mation on our products Stay healthy Microlife AG This device uses the same measuring technology as the award winning BP 3BTO A model tested according to the British and Irish Hypertension Society BIHS protocol Table of Contents 1 Using the Device for the First Time Inserting the batteries Setting the date and time Selecting the correct cuff Selecting the user Selecting standard or MAM mode 2 ...

Page 4: ...t the minutes and pressed the MAM button the date and time are set and the time is displayed 5 If you want to change the date and time press and hold the MAM button for approx 3 seconds until the year number starts to flash Now you can enter the new values as described above Selecting the correct cuff Microlife offers different cuff sizes Select the cuff size to match the circumference of your upp...

Page 5: ...The cuff will now pump up automatically Relax do not move and do not tense your arm muscles until the measurement result is displayed Breathe normally and do not talk 4 The cuff fit check AR on the display indicates that the cuff is perfectly placed If the icon BT A appears the cuff is fitted suboptimally but it is still ok to measure 5 When the correct pressure is reached the pumping stops and th...

Page 6: ...elect either user 1 or 2 by pressing the user button 5 Viewing the average of the last 28 days Press the M button 3 briefly when the device is switched off The display first shows M BP and 28A which stands for the average of measurement values of the last 28 days Blood pressure readings with suboptimal cuff fit BT A are not considered in the average value Viewing the clinical blood pressure averag...

Page 7: ...alues although date and time must be reset the year number therefore flashes automatically after the batteries are replaced Which batteries and which procedure Use 4 new long life 1 5 V size AA alkaline batteries Do not use batteries beyond their date of expiry Remove batteries if the device is not going to be used for a prolonged period Using rechargeable batteries You can also operate this devic...

Page 8: ...e See also chapter Appearance of the Irregular Heartbeat IHB Symbol of this user manual The measurement results given by this device is not a diag nosis It is not replacing the need for the consultation of a physi cian especially if not matching the patient s symptoms Do not rely on the measurement result only always consider other potentially occurring symptoms and the patient s feedback Calling ...

Page 9: ...over in a tumble dryer WARNING Under no circumstances may you wash the inner bladder Accuracy test We recommend this device is tested for accuracy every 2 years or after mechanical impact e g being dropped Please contact your local Microlife Service to arrange the test see foreword Disposal Batteries and electronic devices must be disposed of in accordance with the locally applicable regulations n...

Page 10: ...s reserved The Bluetooth word mark and logos are registered trademarks owned by Bluetooth SIG Inc and any use of such marks by Micro life Corp is under license Other trademarks and trade names are those of their respective owners Expected service life Device 5 years or 10000 measurements Accessories 2 years ...

Reviews: