«AUSCULTATION» Mode
«
AUSCULTATION
» Mode is used for blood pressure
measurement of patients to confirm if a patient is suitable
for the oscillometric method.
Press and hold «
Auscult.
» button to start pumping. Release
the button when it reaches the needed pressure. Press and
hold the «
Auscult.
» button to repump during deflaing; press
«
Start/Stop
» button to release pressure during deflating.
There is no measurement readings displayed in this mode.
The user has to use with a stethoscope to determine the
blood pressure.
«
Cycling
»
SYS
mmHg
DIA
mmHg
PUL
/min
Auscult.
Single Cycling
MAP
Start/Stop
Time/Test
M
(3, Afib)
BASIC
+
*
The device will automatically return to ‘Single’ or ‘Cycling’
mode if there is no operation for one minute after the
auscultation operation.
«Cycling» Mode
(automatic triple measurements)
Select «
Cycling
» Mode then press
«
Start/Stop
» button
to perform automatic triple blood pressure measurement
with 60 seconds intervals. The average reading
measurement readings is displayed and saved after the
measurements.
*
User can set measurement intervals at 15 or 60 seconds in
Cycling Mode. Press «Start/Stop» Button for 3 seconds then
press «Auscult.» Button to select 15 or 60. Press «Start/Stop»
Button to confirm the selection.