BP W70

EN

4. The cuff fit check 

AO

 on the display indicates that the cuff is 

perfectly placed. If the icon 

AO

-A appears, the cuff is fitted subop-

timally, but it is still ok to measure.

5. The measurement is performed during the inflation. The inflation 

speed may vary, this is a normal occurence.

6. During the measurement, the pulse indicator 

AP

 flashes in the display.

7. The result, comprising the systolic 

7

 and the diastolic 

8

 blood 

pressure and the pulse rate 

9

 is displayed. Note also the expla-

nations on further display symbols in this booklet.

8. Remove and switch off the monitor. (The monitor does switch off

auto-matically after approx. 1 min.).

You can stop the measurement at any time by pressing the 

ON/OFF button (e.g. if you feel uneasy or an unpleasant pres-

sure sensation). 

How not to store a reading

As soon as the reading is displayed press and hold the ON/OFF 

button 

1

 until «

M

» 

AK

 is flashing. Confirm to delete the reading by 

pressing the time button 

4

.

«

CL

» is displayed when the reading is deleted from the 

memory successfully.

How do I evaluate my blood pressure?

The triangle on the left-hand edge of the display 

AL

 points at the range 

within which the measured blood pressure value lies. The value is either 

within the optimum (green), elevated (yellow) or high (red) range. 

The classification corresponds to the following ranges defined by 

international guidelines (ESH, ESC, JSH). Data in mmHg.

The higher value is the one that determines the evaluation. Example: 

a blood pressure value of 

140/80

 mmHg or a value of 

130/90

 mmHg 

indicates «blood pressure too high».

Appearance of the Pulse Arrhythmia (PAD)

This symbol 

AM

 indicates that certain pulse irregularities were 

detected during the measurement. In this case, the result may 

deviate from your normal blood pressure – repeat the measurement. 

In most cases, this is no cause for concern. However, if the symbol 

appears on a regular basis (e.g. several times a week with measure-

ments taken daily) we advise you to tell your doctor. Please show 

your doctor the following explanation:

Range

Systolic Diastolic Recommendation 

1. blood pressure 

too high

135

85

Seek medical 

advice

2. blood pressure 

elevated

130 - 134 80 - 84

Self-check

3. blood pressure 

normal

<130

<80

Self-check

is displayed. Breathe normally and do not talk.

Summary of Contents for BP W70

Page 1: ...ia Microlife Corporation Tel 886 2 8797 1288 Fax 886 2 8797 1283 Email service microlife com tw www microlife com North Central South America Microlife USA Inc 1617 Gulf to Bay Blvd 2nd Floor SuiteA Clearwater FL33755 USA Tel 1 727 442 5353 Fax 1 727 442 5377 Email msa microlifeusa com www microlife com IB BP W70 EN ROA 0922 BPW70 Wrist Blood Pressure Monitor EN ...

Page 2: ...Microlife BP W70 7 8 9 AT AK AL AM AN AO AO A AO B AO C AQ AP Microlife BP W70 Guarantee Card Name of Purchaser Serial Number Date of Purchase Specialist Dealer 1 3 5 6 4 2 BT AR AS ...

Page 3: ... an adult individual Dear Customer This device was developed in collaboration with physicians and clinical tests carried out prove its measurement accuracy to be of a very high standard If you have any questions problems or want to order spare parts please contact your local Microlife Customer Service Your dealer or pharmacy will be able to give you the address of the Microlife dealer in your coun...

Page 4: ...est and Disposal Safety and protection Device care Cleaning the cuff Accuracy test Disposal 8 9 Guarantee Technical Specifications 1 Using the Device for the First Time Inserting the batteries After you have unpacked your device first insert the batteries The battery compartment 5 is on the bottom of the device Insert the batteries 2 x 1 5 V size AAA thereby observing the indicated polarity Select...

Page 5: ...e Button to confirm and then enter the date and time setting There are 5 levels of height detection plus a blue LED indicator to help the user determine if the monitor is at the correct height When the monitor is at higher position relative to your heart 1 or 2 downward arrow s appear to remind you to lower the wrist slowly until the blue LED is lighted with a heart symbol Conversely 1 or 2 upward...

Page 6: ...aking a Reliable Measurement 1 Avoid activity eating or smoking immediately before the measurement 2 Sit down on a back supported chair and relax for 5 minutes Keep the feet flat on the floor and do not cross your legs Adjust your seat height to make sure the distance between your seat and the top of the table is within 30 5 cm 12 2 inches 3 Always measure in a sitting position and on the same wri...

Page 7: ...ssfully How do I evaluate my blood pressure The triangle on the left hand edge of the display AL points at the range within which the measured blood pressure value lies The value is either within the optimum green elevated yellow or high red range The classification corresponds to the following ranges defined by international guidelines ESH ESC JSH Data in mmHg The higher value is the one that det...

Page 8: ... daily we recommend the patient toseek medical advice This device does not replace a cardiac examination but serves to detect pulse irregularities at an early stage 6 Memory full Pay attention that the maximum memory capacity of 99 memo ries is not exceeded When the 99 memory is full the oldest value is automatically overwritten with the 100th value Values should be evaluated by a doctor before th...

Page 9: ... batteries and which procedure Use 2 new long life 1 5 V size AAA alkaline batteries Do not use batteries beyond their date of expiry Remove batteries if the device is not going to be used for a prolonged period Using rechargeable batteries You can also operate this device using rechargeable batteries Only use NiMH type reusable batteries Batteries must be removed and recharged when the flat batte...

Page 10: ...is too high over 299 mmHg OR the pulse is too high over 199 beats per minute Relax for 5 minutes and repeat the measurement LO Pulse too low The pulse is too low less than 40 beats per minute Repeat the measurement Error Description Potential cause and remedy 8 ERR 5 Abnormal result The measuring signals are inaccurate and no resultcanthereforebedisplayed Readthrough the checklist for performing r...

Page 11: ...ditional safety information provided within the indi vidual sections of this instruction manual The measurement results given by this device is not a diagnosis It is not replacing the need for the consultation of a physician especially if not matching the patient s symptoms Do not rely on the measurement result only always consider other potentially occurring symptoms and the patient s feedback Ca...

Page 12: ... Battery lifetime IP Class Reference to standards 0 299 mmHg 1 mmHg pressure within 3 mmHg 5 of the readout value 2 x 1 5 V alkaline batteries size AAA approx 300 measurements using new batteries IEC 80601 2 30 ISO 81060 2 IEC 60601 1 IEC 60601 1 2 EMC IEC 60601 1 11 Expected service life Device 5 years or 10000 measurements Accessories 2 years 10 after mechanical impact e g being dropped Please c...

Reviews: