7. Error Messages

If an error occurs during the measurement, the measurement is inter-

rupted and an error message, e.g. «

ERR 3

», is displayed.

Error

Description Potential cause and remedy

«

ERR 1

»

AN

Signal too 

weak

The pulse signals on the cuff are too weak. Re-

position the cuff and repeat the measurement.*

«

ERR 2

»

AO

Error signal During the measurement, error signals were 

detected by the cuff, caused for instance by 

movement or muscle tension. Repeat the 

measurement, keeping your arm still.

«

ERR 3

»

AO

No pressure 

in the cuff

An adequate pressure cannot be generated in 

the cuff. A leak may have occurred. Check that 

the cuff is correctly connected and is not too 

loose. Replace the batteries if necessary. 

Repeat the measurement.

«

HI

»

Pulse or cuff 

«

ERR 6

» MAM Mode

error

«

ERR 

21

»

Wrist Position

Indicator error

pressure too 

high

The pressure in the cuff is too high (over 

299 mmHg) OR the pulse is too high (over 

199 beats per minute). Relax for 5 minutes 

and repeat the measurement.*

«

LO

»

Pulse too 

low

The pulse is too low (less than 40 beats per 

minute). Repeat the measurement.*

Error

Description Potential cause and remedy

8

«

ERR 5

» Abnormal 

result

The measuring signals are inaccurate and no 

result can therefore be displayed. Read through 

the checklist for performing reliable measure-

ments and then repeat the measurement.*

There were too many errors during the 

measurement in MAM mode, making it 

impossible to obtain a final result. Read 

through the checklist for performing reliable 

measurements and then repeat the 

measurement.*

Possibly some problem occurs on the sensor 

for Wrist Position Indicator. Turn off and 

remove the battery then repeat measurement. 

If still displayed, please contact your local 

Microlife-Service.

* Please immediately consult your doctor, if this or any other 

  problem occurs repeatedly.

If you think the results are unusual, please read through the 

information in «Section 1.» carefully.

Summary of Contents for BP W70

Page 1: ...ia Microlife Corporation Tel 886 2 8797 1288 Fax 886 2 8797 1283 Email service microlife com tw www microlife com North Central South America Microlife USA Inc 1617 Gulf to Bay Blvd 2nd Floor SuiteA Clearwater FL33755 USA Tel 1 727 442 5353 Fax 1 727 442 5377 Email msa microlifeusa com www microlife com IB BP W70 EN ROA 0922 BPW70 Wrist Blood Pressure Monitor EN ...

Page 2: ...Microlife BP W70 7 8 9 AT AK AL AM AN AO AO A AO B AO C AQ AP Microlife BP W70 Guarantee Card Name of Purchaser Serial Number Date of Purchase Specialist Dealer 1 3 5 6 4 2 BT AR AS ...

Page 3: ... an adult individual Dear Customer This device was developed in collaboration with physicians and clinical tests carried out prove its measurement accuracy to be of a very high standard If you have any questions problems or want to order spare parts please contact your local Microlife Customer Service Your dealer or pharmacy will be able to give you the address of the Microlife dealer in your coun...

Page 4: ...est and Disposal Safety and protection Device care Cleaning the cuff Accuracy test Disposal 8 9 Guarantee Technical Specifications 1 Using the Device for the First Time Inserting the batteries After you have unpacked your device first insert the batteries The battery compartment 5 is on the bottom of the device Insert the batteries 2 x 1 5 V size AAA thereby observing the indicated polarity Select...

Page 5: ...e Button to confirm and then enter the date and time setting There are 5 levels of height detection plus a blue LED indicator to help the user determine if the monitor is at the correct height When the monitor is at higher position relative to your heart 1 or 2 downward arrow s appear to remind you to lower the wrist slowly until the blue LED is lighted with a heart symbol Conversely 1 or 2 upward...

Page 6: ...aking a Reliable Measurement 1 Avoid activity eating or smoking immediately before the measurement 2 Sit down on a back supported chair and relax for 5 minutes Keep the feet flat on the floor and do not cross your legs Adjust your seat height to make sure the distance between your seat and the top of the table is within 30 5 cm 12 2 inches 3 Always measure in a sitting position and on the same wri...

Page 7: ...ssfully How do I evaluate my blood pressure The triangle on the left hand edge of the display AL points at the range within which the measured blood pressure value lies The value is either within the optimum green elevated yellow or high red range The classification corresponds to the following ranges defined by international guidelines ESH ESC JSH Data in mmHg The higher value is the one that det...

Page 8: ... daily we recommend the patient toseek medical advice This device does not replace a cardiac examination but serves to detect pulse irregularities at an early stage 6 Memory full Pay attention that the maximum memory capacity of 99 memo ries is not exceeded When the 99 memory is full the oldest value is automatically overwritten with the 100th value Values should be evaluated by a doctor before th...

Page 9: ... batteries and which procedure Use 2 new long life 1 5 V size AAA alkaline batteries Do not use batteries beyond their date of expiry Remove batteries if the device is not going to be used for a prolonged period Using rechargeable batteries You can also operate this device using rechargeable batteries Only use NiMH type reusable batteries Batteries must be removed and recharged when the flat batte...

Page 10: ...is too high over 299 mmHg OR the pulse is too high over 199 beats per minute Relax for 5 minutes and repeat the measurement LO Pulse too low The pulse is too low less than 40 beats per minute Repeat the measurement Error Description Potential cause and remedy 8 ERR 5 Abnormal result The measuring signals are inaccurate and no resultcanthereforebedisplayed Readthrough the checklist for performing r...

Page 11: ...ditional safety information provided within the indi vidual sections of this instruction manual The measurement results given by this device is not a diagnosis It is not replacing the need for the consultation of a physician especially if not matching the patient s symptoms Do not rely on the measurement result only always consider other potentially occurring symptoms and the patient s feedback Ca...

Page 12: ... Battery lifetime IP Class Reference to standards 0 299 mmHg 1 mmHg pressure within 3 mmHg 5 of the readout value 2 x 1 5 V alkaline batteries size AAA approx 300 measurements using new batteries IEC 80601 2 30 ISO 81060 2 IEC 60601 1 IEC 60601 1 2 EMC IEC 60601 1 11 Expected service life Device 5 years or 10000 measurements Accessories 2 years 10 after mechanical impact e g being dropped Please c...

Reviews: