10
website:
www.microlife.com/support
Compensation is limited to the value of the product. The guarantee
will be granted if the complete product is returned with the original
invoice. Repair or replacement within guarantee does not prolong
or renew the guarantee period. The legal claims and rights of
consumers are not limited by this guarantee.
13.Technical specifications
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.
Operating condi
-
tions:
10 - 40 °C / 50 - 104 °F
15 - 90 % relative maximum humidity
Storage conditions:
-20 - +55 °C / -4 - +131 °F
15 - 90 % relative maximum humidity
Weight:
393 g (including batteries)
Dimensions:
152 x 92 x 42 mm
Cuff size:
from 17 to 52 cm according to the cuff
sizes (see «Selecting the correct cuff»)
Measuring proce
-
dure:
oscillometric, corresponding to Korotkoff
method: Phase I systolic, Phase V
diastolic
Measurement range:
20 - 280 mmHg – blood pressure
40 - 200 beats per minute – pulse
Cuff pressure display
range:
0 - 299 mmHg
Resolution:
1 mmHg
Static accuracy:
within ± 3 mmHg
Pulse accuracy:
± 5 % of the readout value
Voltage source:
4 x 1.5 V alkaline batteries; size AA
Mains adapter DC 6V, 600 mA
(optional)
Battery lifetime:
approx. 920 measurements
(using new batteries)
IP Class:
IP20
Reference to
standards:
EN 1060-1 /-3 /-4; IEC 60601-1;
IEC 60601-1-2 (EMC); IEC 60601-1-11
Expected service life:
Device: 5 years or 10000 measurements,
whichever comes first
Accessories: 2 years or 5000 measure
-
ments, whichever comes first
Summary of Contents for BP A150 AFIB
Page 18: ...16 6 1 7 8 9 AS 10 AL AM AN 11 12 1 y 1 M AR M 2 5 AF BT 6 BM 30 mmHg 40 mmHg AFIB AFIB AFIB...
Page 21: ...19 Microlife BP A150 AFIB RU 12...
Page 22: ...20 5 ME EN 60601 1 2 2015...
Page 33: ...31 Microlife BP A150 AFIB BG 12...