5
Pressing the M-button again displays the previous value. Pressing
the M-button repeatedly enables you to move from one stored value
to another.
Memory full
Pay attention that the maximum memory capacity of 100 is
not exceeded.
When the memory is full, the old values
are automatically overwritten with new ones.
Values
should be evaluated by a doctor before the memory capacity
is reached – otherwise data will be lost.
Clearing all values
If you are sure that you want to permanently remove all stored
values, hold down the M-button (the instrument must have been
switched off beforehand) until «
CL
» appears and then release the
button. To permanently clear the memory, press the M-button while
«
CL
» is flashing. Individual values cannot be cleared.
How not to store a reading
Press the ON/OFF button (1) while the reading is being displayed.
Keep the button pressed until «
M
» (16) is flashing and then release
it. Confirm by pressing the M-button.
7. Setting the Alarm Function
This instrument allows you to set 2 alarm times at which an alarm
signal will then be triggered. This can be a useful aid, for instance
as a reminder to take medication.
1. To set an alarm time, press the time button (11) (the instrument
must have been switched off beforehand) and immediately
afterwards the M-button (10) and hold both down until the
bell symbol (21) appears in the bottom left of the display. Then
release both buttons. The flashing «
1
» in the display indicates
that the first alarm time can now be set.
2. Press the time button to set the hours – the hour display flashes
and pressing the M-button allows you to set the alarm hour. To
confirm, press the time button.
3. The minute display will now flash. The minutes can be set using
the M-button. To confirm, press the time button again.
4.
The bell symbol will now flash. Use the M-button to select
whether the alarm time is to be active (bell) or inactive
(crossedout bell). To confirm, press the time button.
•
To set a second alarm time, proceed as above but if the «
1
»
flashes, press the M-button to select «
2
» and confirm with the
time button.
•
An active alarm time is indicated by the bell symbol in the
display.
•
The alarm will sound at the set time every day.
•
To switch-off the alarm when it is sounding, press the ON/OFF
button (1).
•
To permanently switch off the alarm, proceed as above and
select the crossed-out bell symbol. This will then disappear from
the display.
•
The alarm times must be re-entered each time the batteries are
replaced.
8. Battery Indicator and Battery change
Low battery
When the batteries are approximately 3/4 used, the battery symbol
(18) will flash as soon as the instrument is switched on (partly
filled battery displayed). Although the instrument can still measure
reliably, you should obtain replacement batteries.
Flat battery - replacement
When the batteries are flat, the battery symbol (18) will flash as
soon as the instrument is switched on (flat battery is displayed).
You cannot take any further measurements and must replace the
batteries.
1. Open the battery compartment (7) at the back of the instrument.
2. Replace the batteries - ensure correct polarity as shown by the
symbols in the compartment.
3. To set date and time, follow the procedure described in «Section 2.».
➢
The memory retains all values although date and time (and
possibly also set alarm times) must be reset – the year number
therefore flashes automatically after the batteries are replaced.
Which batteries and which procedure?
➢
Please use 4 new, long-life 1.5V, size AA batteries.
➢
Do not use batteries beyond their date of expiry.
➢
Remove batteries, if the instrument is not going to be used for a
prolonged period.