background image

4

Checklist for taking a reliable measurement

1.  Avoid activity, eating or smoking immediately before the 

measurement.

2.  Sit down for at least 5 minutes before the measurement  and 

relax.

3.  Always measure on the same arm (normally left).
4.  Remove close-fitting garments from the upper arm. To avoid 

constriction, shirt sleeves should not be rolled up - they do not 
interfere with the cuff if they are laid flat.

5.  Always ensure that the cuff is positioned correctly, as shown on 

the cuff.

      

• Fit the cuff closely, but not too tight.

      

• Make sure that the cuff is 2 cm (0.75 inch) above your elbow  

          with the tube on the inside of your arm.
      

• Support your arm so it is relaxed.

      

• Ensure that the cuff is at the same height as your heart.

6.  Press the ON/OFF button (1) to start the measurement.
7.  The cuff will now pump up automatically. Relax, do not move 

and do not tense your arm muscles until the measurement result 
is displayed. Breathe normally and do not talk.

8.  When the correct pressure is reached, the pumping stops and 

the pressure falls gradually. If the required pressure was not 
reached, the instrument will automatically pump some more air 
to the cuff.

9.  During the measurement, the heart symbol (19) flashes in the 

display and a beep sounds every time a heartbeat is detected.

10. The result, comprising the systolic (13) and the diastolic (14)

blood pressure and the pulse (15) is displayed and longer beep 
is heard. At the same time, the blood pressure monitor will 
announce the result. Note also the explanations on further 
displays in this booklet.

11. When the measurement is finished, remove the cuff.
12. Switch off the instrument. (The monitor does switch off 

automatically after approx. 3 min.).

  You can stop the measurement at any time by pressing the ON/

OFF button (e.g. if you feel uneasy or an unpleasant pressure 
sensation).

4.  Appearance of the Heart Arrhythmia Indicator for

     early Detection

This symbol (17) indicates that certain pulse irregularities were 
detected during the measurement. In this case, the result may 
deviate from your normal blood pressure – repeat the measurement.
In most cases, this is no cause for concern. However, if the 
symbol appears on a regular basis (e.g. several times a week with 
measurements taken daily) we advise you to tell your doctor.
Please show your doctor the following explanation:

Information for the doctor on frequent appearance of the
Arrhythmia indicator

This instrument is an oscillometric blood pressure monitor that also 
analyses pulse frequency during measurement. The instrument is 
clinically tested.
The arrhythmia symbol is displayed after the measurement, if pulse 
irregularities occur during measurement. If the symbol appears more 
frequently (e.g. several times per week on measurements performed 
daily) we recommend the patient to seek medical advice.
The instrument does not replace a cardiac examination, but serves 
to detect pulse irregularities at an early stage.

5.  Traffic Light Indication in the Display

The bars on the left-hand edge of the traffic light display (20) show 
you the range within which the indicated blood pressure value lies.
Depending on the height of the bar, the readout value is either within 
the  normal  (green),  borderline  (yellow)  or  danger  (red)  range.  The 
classification corresponds to the 6 ranges in the Table as defined by 
the WHO, as described in «Section 1.».

6.  Data Memory

At the end of each measurement, this instrument automatically
stores the result, including date and time.

Viewing the stored values

Press the M-button (10) briefly, when the instrument is switched
off. The display first shows «

M

» (16) and then a value, e.g. «

M 17

».

This means that there are 17 values in the memory. The instrument
then switches to the last stored result.

Summary of Contents for BP 3AP1-3E

Page 1: ...3 Microlife BP 3AP1 3E Automatic Blood Pressure Monitor Instruction Manual...

Page 2: ...4 Microlife BP 3AP1 3E...

Page 3: ...nstrument was developed in collaboration with physicians clinical tests have proven its measurement accuracy to be very high Please read through these instructions carefully so that you understand all...

Page 4: ...ues can damage your health and must be treated by your doctor Always discuss your values with your doctor and tell him her if you have noticed anything unusual or feel unsure Never rely on single bloo...

Page 5: ...2 You can now set the month using the M button Press the time button to confirm and then set the day 3 Please follow the instructions above to set the day hour and minutes 4 Once you have set the min...

Page 6: ...n e g if you feel uneasy or an unpleasant pressure sensation 4 Appearance of the Heart Arrhythmia Indicator for early Detection This symbol 17 indicates that certain pulse irregularities were detected...

Page 7: ...ng the M button To confirm press the time button again 4 The bell symbol will now flash Use the M button to select whether the alarm time is to be active bell or inactive crossedout bell To confirm pr...

Page 8: ...list for performing reliable measurements and then repeat the measurement HI Pulse or cuff The pressure in the cuff is too high over pressure too 300 mmHg or the pulse is too high over high 200 beats...

Page 9: ...rating temperature 10 40 50 104 Storage temperature 20 55 4 131 15 90 relative maximum humidity Weight 480 g including batteries Dimensions 180 x 110 x 75 mm Measuring procedure oscillometric correspo...

Reviews: